ThermoJacket
Users Manual, Rev. 1.1, Jan 2019
28
Sequence to extract the sensor out of the ThermoJacket
•
Open the snap locks, release the rear ThermoJacket cover and leave the rear cover hang down,
supported by flexible wire rope.
•
Unscrew all top cable grommet compression seal nuts and push them downwards along the cables.
Use a spanner with 30 mm (1.2 in) wrench size to release the sealed cables.
•
Unscrew all cable connectors from the Endurance sensor by hand power and pull the cables completely
out of the ThermoJacket grommet.
•
Turn the Endurance sensor by hand power, induced to one of both cable sockets, some degrees anti-
clockwise to give free space for the afterwards attached extraction tool.
•
Press the flexible rear cover wire rope and the attached screw eye with your left-hand thumb to the left
ThermoJacket inside.
•
Screw the extraction tool by right hand power onto the left M16 Endurance sensor socket. Use the
whole thread pitch to fix the tool on the socket.
•
Keep the flexible wire rope and the screw eye in the outmost left position by your left-hand thumb.
•
Force to turn the Endurance sensor via the installed extraction tool in clockwise rotating direction, to
induce an ejection force over a die cast metal wedge.
•
Pull the released Endurance sensor easily by hand power out of the ThermoJacket.
1.800.561.8187