6  Fluke Corporation     ScopeMeter 190 Series II

190-062

190-102

190-202

190-502

190-104

190-204

Input voltage range

Rated maximum floating voltage CAT III 1000 V/CAT IV 600 V (maximum voltage between any contact and earth-ground voltage level)
Probe input voltage VPS410

CAT III 1000 V/CAT IV 600 V (Maximum voltage between 10:1 probe tip and reference lead)

Probe input voltage VPS510

CAT III 300 V (Maximum voltage between 10:1 probe tip and reference lead) 

Maximum BNC input voltage

CAT IV 300 V (maximum voltage on BNC input directly)

Maximum voltage  

on meter input

CAT III 1000 V/CAT IV 600 V  

(safety designed banana input connectors)

Memory save and recall

Memory locations (internal)

30 waveform memories plus 10 recording memories plus 9 screen copy memories (190-XX, 2 channel  

models); 15 waveforms memories plus 2 recording memories plus 1 screen copy memory (190-XX, 4 channel 

models)

15 waveform memory locations Stores Scope-trace waveform data (2 or 4 traces each) plus screen-copy plus corresponding setup
Two recording memories

Each may contain: 

  a 100 Screen Replay sequence, or

  a ScopeRecord Roll-mode recording (2 or 4 traces), or

  a TrendPlot recording of up to 4 measurements

External data storage

  On PC, using FlukeView

 Software, or

  Direct storage on external flash memory drive (maximum 2 GB) through USB host port

Screencopies

  On PC, using FlukeView

 Software, or

  Internally (in instrument) which can be copied on to external flash memory drive as .BMP-file,  

through USB host port

Volatility

Measurement data is initially stored in RAM, which is maintained by the main battery with a 30 seconds 

back-up when battery is exchanged 

When storing data, this is written in non-volatile flash-ROM

Real-time clock

Provides date and time stamp information for ScopeRecord, for 100 Screen Replay sequences and  

for TrendPlot recordings

Case

Design

Rugged, shock-proof with integrated protective holster. Handstrap and hangstrap included as standard 

Kensington lock supported to lock down instrument when left unattended

Drip and dust proof

IP 51 according to IEC529

Shock and vibration

Shock 30 g, vibration (sinusoidal) 3 g according to MIL-PRF-28800F Class 2

Display size

127 mm x 88 mm (153 mm/6.0 in diagonal) LCD

Resolution

320 x 240 pixels

Contrast and brightness

User adjustable, temperature compensated

Brightness

200 cd/m

2

  typ. using power adapter, 90 cd/m

2

  typical using battery power

Mechanical data

Size

265 mm x 190 mm x 70 mm (10.4 in x 7.5 in x 2.8 in)

Weight (including battery)

2.1 kg (4.6 lb)

2.2 kg (4.8 lb)

Power

Line power

Mains adapter/battery charger BC190 included, version depending of country

Battery power

Re-chargeable double capacity Li-Ion battery (included). Battery swappable through  

easily accessible battery door at the rear of the instrument

Battery type (incl.) and capacity 

[+opt. battery]

BP290; 2400 mAh  

[BP291 (4800 mAh) optional]

BP291; 4800 mAh

Battery charge indicator

Battery has built-in status indicator for use with external charger,  

next to battery status indicator on instrument screen

Battery operating time  

(with backlight low)

Up to four hours using BP290 (included), 

Up to eight hours using BP291 (optional)

Up to seven hours using BP291 (included)

Battery charging time

2½ hours using BP290; 5 hours using BP291

Five hours BP291

Battery power saving functions

Auto ‘power down’ with adjustable power down time; Auto ‘Display off’ with adjustable power down time; 

On-screen battery power indicator

Safety

Compliance

EN61010-1-2001, Pollution Degree 2;  

CAN/CSA C22.2, No. 61010-1-04, with approval; UL61010B; ANSI/ISA-82.02.01

General Specifications

Shop for Fluke products online at:

1.877.766.5412

www.

MyFlukeStore

.com

Содержание ScopeMeter 190-062

Страница 1: ...nect and View triggering for intelligent automatic triggering on fast slow and even complex signals Frequency Spectrum using FFT analysis Automatic capture and REPLAY of 100 screens ScopeRecord Roll mode gives 30 000 points per input channel for low frequency signal analysis TrendPlot paperless recorder mode with deep memory for long term automatic measurements 5 000 count DMM included in the 2 ch...

Страница 2: ... 8 ns peak detect on each channel using real time sampling and data compression at any timebase setting Display and acquisition Display 153 mm 6 in full color LCD with LED backlight Display modes Any combination of channels average on off replay Visible screen width 12 divisions horizontally in scope mode Digital persistence modes off short medium long infinite and envelope mode Waveform mathemati...

Страница 3: ...ss Fail Testing In waveform compare mode the oscilloscope can be set up to store only matching Pass or only non matching Fail acquired waveforms in the replay memory bank for further analysis Automatic scope measurements V dc V ac rms V ac dc Vpeak max Vpeak min Vpeak to peak A ac A dc A ac dc frequency in Hz risetime using cursors falltime using cursors Power Factor PF Watts VA VA reactive phase ...

Страница 4: ...28 C Voltage V dc accuracy 0 5 5 counts 1 5 5 counts V ac true rms accuracy 1 10 counts 2 5 15 counts 15 Hz to 60 Hz 1 5 10 counts 60 Hz to 1 kHz 60 Hz to 20 kHz 2 5 15 counts V ac dc true rms accuracy 1 10 counts 2 5 15 counts 15 Hz to 60 Hz 1 5 10 counts 60 Hz to 1 kHz 60 Hz to 20 kHz 2 5 15 counts Voltmeter ranges 500 mV 5 V 50 V 500 V 1 000 V Resistance Ranges 500 Ω 5 kΩ 50 kΩ 500 kΩ 5 MΩ 30 M...

Страница 5: ...imespan 6 sec 48 hr Time division in view all mode 0 5 s div 4 h div Glitch capture 8 ns Sample rate 125 MS s Resolution 200 μsec 4 8 sec Trendplot Recording Multiple channel electronic paperless recorder Graphically plots displays and stores results of up to four automatic scope measurements or a DMM reading over time Source and display Any combination of scope measurements made on any of the inp...

Страница 6: ...vides date and time stamp information for ScopeRecord for 100 Screen Replay sequences and for TrendPlot recordings Case Design Rugged shock proof with integrated protective holster Handstrap and hangstrap included as standard Kensington lock supported to lock down instrument when left unattended Drip and dust proof IP 51 according to IEC529 Shock and vibration Shock 30 g vibration sinusoidal 3 g a...

Страница 7: ...l and data transfer under PC control Probe calibration output Dedicated probe cal output with reference contact provided fully insulated from any measurement input channel Warranty Three years parts and labor on main instrument one year on accessories Included accessories Battery charger mains adapter BC190 Li Ion battery pack BP290 2400 mAh BP291 4800 mAh Voltage probe sets Each set includes grou...

Страница 8: ...nic Voltage Probe set 10 1 500 MHz one set green VPS410 R Industrial Voltage Probe set 10 1 one set red VPS410 G Industrial Voltage Probe set 10 1 one set grey VPS410 B Industrial Voltage Probe set 10 1 one set blue VPS410 V Industrial Voltage Probe set 10 1 one set green VPS420 R High working voltage ruggedized probe set 100 1 150 MHz bicolored red black SW90W FlukeView ScopeMeter Software packag...

Отзывы: