Fluke ScopeMeter 190-062 Скачать руководство пользователя страница 3

ScopeMeter 190 Series II      Fluke Corporation  3

Automatic capture of 100 screens

When in oscilloscope mode, the instrument ALWAYS memorizes the last 100 screens—no specific user setup required. When an anomaly is 

seen, the REPLAY button can be pressed to review the full sequence of screen events over and over. Instrument can be set up for triggering on 

glitches or intermittent anomalies and will operate in “baby-sit” mode capturing 100 specified events
Replay

Manual or continuous replay. Displays the captured 100 screens as a “live” animation, or under manual 

control. Each screen has date and time-stamp.

Replay storage

Two sets of 100 screens each can be saved internally for later recall and analysis. 

Direct storage of additional sets on external flash memory drive through USB host port.

FFT—frequency spectrum analysis

Shows frequency content of oscilloscope waveform using Fast Fourier Transform
Window

Automatic, Hamming, Hanning or None

Automatic window

Digitally re-samples acquired waveform to get optimum frequency resolution in FFT resultant

Vertical scale

Linear/Logarithmic (in volts or amps)

Frequency axis

Frequency range automatically set as a function of timebase range of oscilloscope

Waveform compare and pass/fail testing

Waveform Compare

Provides storage and display of a reference waveform for visual comparison with newly acquired  

waveforms. Reference is derived from an acquired waveform and can be modified in the oscilloscope

Pass/Fail Testing 

In waveform compare mode, the oscilloscope can be set up to store only matching (“Pass”) or only  

non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis

Automatic scope measurements

V dc, V ac rms, V ac+dc, Vpeak max, Vpeak min, Vpeak to peak, A ac, A dc, A ac+dc, frequency (in Hz), risetime (using cursors), falltime (using 

cursors), Power Factor (PF), Watts, VA, VA reactive, phase (between any 2 inputs), pulsewidth (pos./neg.), dutycycle (pos./neg.), temperature °C, 

temperature °F (not for Japan), dBV, dBm into 50 I and 600 I, V

PWM

 ac and V

PWM

(ac+dc) for measurement on pulsewidth modulated motordrives 

and frequency inverters, V/Hz ration (190-xx2 only)
Advanced power and motor drive 

functions

V/Hz ratio (190-x02 only), Power Factor (PF), Watts, VA, VA reactive, V

PWM

ac and V

PWM 

(ac+dc) for 

measurement on pulsewidth modulated motordrives and frequency inverters

Advanced functions

mA*s (current-over-time, between cursors);  

V*s (voltage over time, between cursors);  

W*s (energy, between cursors)

Cursor measurements

Source

On any input waveform or on mathematical resultant waveform (excl. X-Y-mode)

Dual horizontal lines

Voltage at cursor 1 and at cursor 2, voltage between cursors

Dual vertical lines

Time between cursors, 1/T between cursors (in Hz), voltage between markers, risetime with markers, 

falltime with markers; Vrms between cursors, Watts between cursors

Single vertical line

Min-Max and Average voltage at cursor position; frequency and rms-value of individual frequency 

component in the FFT Resultant

ZOOM

Ranges from full record overview to zoom in up to sample level, at any record length

Shop for Fluke products online at:

1.877.766.5412

www.

MyFlukeStore

.com

Содержание ScopeMeter 190-062

Страница 1: ...nect and View triggering for intelligent automatic triggering on fast slow and even complex signals Frequency Spectrum using FFT analysis Automatic capture and REPLAY of 100 screens ScopeRecord Roll mode gives 30 000 points per input channel for low frequency signal analysis TrendPlot paperless recorder mode with deep memory for long term automatic measurements 5 000 count DMM included in the 2 ch...

Страница 2: ... 8 ns peak detect on each channel using real time sampling and data compression at any timebase setting Display and acquisition Display 153 mm 6 in full color LCD with LED backlight Display modes Any combination of channels average on off replay Visible screen width 12 divisions horizontally in scope mode Digital persistence modes off short medium long infinite and envelope mode Waveform mathemati...

Страница 3: ...ss Fail Testing In waveform compare mode the oscilloscope can be set up to store only matching Pass or only non matching Fail acquired waveforms in the replay memory bank for further analysis Automatic scope measurements V dc V ac rms V ac dc Vpeak max Vpeak min Vpeak to peak A ac A dc A ac dc frequency in Hz risetime using cursors falltime using cursors Power Factor PF Watts VA VA reactive phase ...

Страница 4: ...28 C Voltage V dc accuracy 0 5 5 counts 1 5 5 counts V ac true rms accuracy 1 10 counts 2 5 15 counts 15 Hz to 60 Hz 1 5 10 counts 60 Hz to 1 kHz 60 Hz to 20 kHz 2 5 15 counts V ac dc true rms accuracy 1 10 counts 2 5 15 counts 15 Hz to 60 Hz 1 5 10 counts 60 Hz to 1 kHz 60 Hz to 20 kHz 2 5 15 counts Voltmeter ranges 500 mV 5 V 50 V 500 V 1 000 V Resistance Ranges 500 Ω 5 kΩ 50 kΩ 500 kΩ 5 MΩ 30 M...

Страница 5: ...imespan 6 sec 48 hr Time division in view all mode 0 5 s div 4 h div Glitch capture 8 ns Sample rate 125 MS s Resolution 200 μsec 4 8 sec Trendplot Recording Multiple channel electronic paperless recorder Graphically plots displays and stores results of up to four automatic scope measurements or a DMM reading over time Source and display Any combination of scope measurements made on any of the inp...

Страница 6: ...vides date and time stamp information for ScopeRecord for 100 Screen Replay sequences and for TrendPlot recordings Case Design Rugged shock proof with integrated protective holster Handstrap and hangstrap included as standard Kensington lock supported to lock down instrument when left unattended Drip and dust proof IP 51 according to IEC529 Shock and vibration Shock 30 g vibration sinusoidal 3 g a...

Страница 7: ...l and data transfer under PC control Probe calibration output Dedicated probe cal output with reference contact provided fully insulated from any measurement input channel Warranty Three years parts and labor on main instrument one year on accessories Included accessories Battery charger mains adapter BC190 Li Ion battery pack BP290 2400 mAh BP291 4800 mAh Voltage probe sets Each set includes grou...

Страница 8: ...nic Voltage Probe set 10 1 500 MHz one set green VPS410 R Industrial Voltage Probe set 10 1 one set red VPS410 G Industrial Voltage Probe set 10 1 one set grey VPS410 B Industrial Voltage Probe set 10 1 one set blue VPS410 V Industrial Voltage Probe set 10 1 one set green VPS420 R High working voltage ruggedized probe set 100 1 150 MHz bicolored red black SW90W FlukeView ScopeMeter Software packag...