8-1
Chapter 8
General Information
Introduction
This instrument employs contemporary time-proved analog techniques. The components
were chosen for their long-term stability and reliability. Numerous test points exist on the
circuit board that facilitate trouble-shooting, when necessary. The operational amplifiers
are plug-in, which allows for easy installation.
The 15-volt DC system is powered by a modular power supply which is, in turn, fully
isolated from line voltages through a dual primary input isolation transformer. The input
line voltage, 120 Vac or 230 Vac at 50 Hz or 60 Hz, is selectable by means of a small
circuit board located internal to the power cord receptacle. Removal of the power cord
allows access to the fuse and circuit board compartment.
The front panel controls include a power switch and indicator, a 3.5-inch null meter, a
ten-turn ZERO potentiometer with calibrated dial, a ten-turn GAIN potentiometer with
calibrated dial, and a SHUNT switch that allows discrete high and low gain selection.