Fluke ProSim 4 Скачать руководство пользователя страница 27

 

Vital Signs Simulator 

 

Maintenance 

15 

Battery Removal 

The battery pack is easy to remove and replace. To 
remove the battery pack: 

1.  Push down on the battery pack latch as shown in 

Figure 5. 

2.  Pull the battery pack from the Product. 

To put the battery pack into the Product, align the battery 
pack with the guides on the Product and push it into the 
Product until the catch locks. 

The ProSim 4 battery is not compatible with the 
ProSim 6/8.

Push

Down

Pull Out

 

gne023.eps 

Figure 5. Battery Removal 

Содержание ProSim 4

Страница 1: ...1 Rev 3 2 15 2011 2015 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies ProSim 4 Vital Signs Sim...

Страница 2: ...ke Biomedical service facility NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMA...

Страница 3: ...sk for an agent to be present while the instrument is unpacked There are no special unpacking instructions but be careful not to damage the instrument when unpacking it Inspect the instrument for phys...

Страница 4: ...at least four inches of tightly packed industry approved shock absorbent material around the instrument Returns for partial refund credit Every product returned for refund credit must be accompanied...

Страница 5: ...tions may result in electrical shock hazards or improper operation Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications Restrictions and Liab...

Страница 6: ......

Страница 7: ...Symbols 3 Unpack the Product 3 Accessories 4 Instrument Familiarization 6 How to Turn On the Product 10 How to Change the Display Language 10 Maintenance 11 How to Clean the Product 12 Battery Mainten...

Страница 8: ...g Started ii Detailed Specifications 17 Normal Sinus Rhythm Waveform 17 Arrhythmia 17 ECG Performance Testing 18 Respiration 18 Invasive Blood Pressure 19 Non Invasive Blood Pressure 19 Presets and Au...

Страница 9: ...iii List of Tables Table Title Page 1 Simulation Types 1 2 Symbols 3 3 Standard Accessories 4 4 Optional Accessories 5 5 Product Controls and Connections 7 6 Display Features 9...

Страница 10: ...ProSim 4 Getting Started iv...

Страница 11: ...v List of Figures Figure Title Page 1 Product Controls and Connections 6 2 Display Features 8 3 Home Screen 10 4 External Battery Charging Connection 14 5 Battery Removal 15...

Страница 12: ...ProSim 4 Getting Started vi...

Страница 13: ...s used to monitor various physiological parameters of a patient including ECG respiration invasive blood pressure and non invasive blood pressure The intended user is a trained biomedical equipment te...

Страница 14: ...t be closed and locked before you operate the Product Remove all probes test leads and accessories that are not necessary for the measurement Do not use the Product around explosive gas vapor or in da...

Страница 15: ...Contact your authorized Fluke Service Center for recycling information This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicates that you must not disca...

Страница 16: ...art Number ProSim 4 Getting Started Manual 3931478 ProSim 4 Users Manual CD 3931519 AC DC Power Supply 3978380 AC Power Cord US 284174 Schuko 769422 UK 769455 Japan 284174 Australia 658641 Brazil 1 38...

Страница 17: ...t Number Battery pack 4026823 USB Cable Mini Series B 1 meter long 4034393 NIBP Mandrel Set 4308086 Modules to convert ECG snap adapter to 4 mm and 3 2 mm ECG banana adapter as part of optional access...

Страница 18: ...ProSim 4 Getting Started 6 Instrument Familiarization Table 5 is a list of Product controls and connections shown in Figure 1 5 1 2 4 6 7 8 9 3 3 gne019 eps Figure 1 Product Controls and Connections...

Страница 19: ...s and calibration ECG Posts Connection posts for Device Under Test DUT ECG leads Air Port Connector Pressure port for NIBP cuff and monitor IBP Channel 1 Connector Connector to an IBP input of the pat...

Страница 20: ...ProSim 4 Getting Started 8 Figure 2 and Table 6 identify the display features 1 2 3 5 4 gne010 eps Figure 2 Display Features...

Страница 21: ...reen name Battery ICON that indicates the charge level of the battery Simulation parameters Shows the simulation parameter values Controls Touch sensitive controls to set simulation parameters and Pro...

Страница 22: ...een All Product simulations and tests are set through the controls on the Home screen How to Change the Display Language To change the language used in the display 1 From the home screen shown in Figu...

Страница 23: ...r batteries in a container where the terminals can be shorted Connect the battery charger to the mains power outlet before the Product Repair the Product before use if the battery leaks Remove batteri...

Страница 24: ...e Product Use only specified replacement parts Have an approved technician repair the Product How to Clean the Product Caution Do not pour fluid onto the Product surface fluid seepage into the electri...

Страница 25: ...shown in the upper right corner of the display Shows when the ac dc power supply is connected Shows the battery level when the Product operates on the battery The battery can be charged while it is i...

Страница 26: ...ProSim 4 Getting Started 14 Battery LED gne022 eps Figure 4 External Battery Charging Connection...

Страница 27: ...he battery pack latch as shown in Figure 5 2 Pull the battery pack from the Product To put the battery pack into the Product align the battery pack with the guides on the Product and push it into the...

Страница 28: ...Weight 0 88 kg 1 93 lb Safety IEC 61010 1 Category II Pollution Degree 2 Electromagnetic Compatibility EMC International IEC 61326 1 Basic Electromagnetic Environment CISPR 11 Group 1 Class A Group 1...

Страница 29: ...rds Amplitude 1 0 mV Other leads are proportional to Lead II reference lead in percentage per Lead I 70 Lead II 100 Lead III 30 Lead V1 24 Lead V2 48 Lead V3 100 Lead V4 120 Lead V5 112 Lead V6 80 Amp...

Страница 30: ...Amplitude 1 mV Other leads are proportional to Lead II reference lead in percentage per Lead I 70 Lead II 100 Lead III 30 Lead V1 24 Lead V2 48 Lead V3 100 Lead V4 120 Lead V5 112 Lead V6 80 Square W...

Страница 31: ...ccuracy guaranteed for DC excitation only Static Pressure 0 80 160 and 250 mmHg Dynamic Waveforms Synchronization To ECG heartrate Chambers simulated systolic diastolic pressure Type IBP Arterial IBP...

Страница 32: ...ndependent of device under test Synchronization To ECG heartrate maximal rate 120 BPM Leak Test Target Pressure 20 to 400 mmHg Elapse time 0 30 to 5 00 minutes seconds in 30 second steps Leakage Rate...

Отзывы: