™/S, MOLBLOC TERMINAL OPERATION AND MAINTENANCE MANUAL
© 1995 - 2010 Fluke Corporation, DH Instruments Division
Page 146
1.
Current
adder
(always in Pascal [Pa]). This is
an edit field if
<3edit>
was selected.
2.
Indication of which RPT calibration information
is being viewed/edited.
3.
This is an Current
multiplier
(dimensionless).
This is an edit field if
<3edit>
was selected.
PA: 0.0 dnstrm RPT
PM: 1.00000
If
<3edit>
was selected, edit the values of
PA
and
PM
as desired.
Pressing
[ESCAPE]
returns to the
<1run 2view 3edit>
screen without
making changes. Press
[ENTER]
to continue.
The display is:
1.
Current calibration date in YYYYMMDD
format for the selected RPT. This is an edit
field if
<3edit>
was selected. The default date
is 19800101..
dnstrm RPT cal date:
19990214
If
<3edit>
was selected, edit the calibration date as desired using a
YYYYMMDD format. Pressing
[ESCAPE]
returns to the
<1run 2view 3edit>
screen without making changes. Press
[ENTER]
to continue. If
<3edit>
was
selected and changes have been made to
PA
,
PM
or the calibration date, a
confirmation screen is presented to
<Activate changes>
. Pressing
[ESCAPE]
or selecting
<1no>
returns to the
<1run 2view 3edit>
screen without
making changes. Selecting
<2yes>
activates changes and returns to the
<1run
2view 3edit>
screen.
Note
•
There is a specific calibration date for each RPT. Changing
the calibration date of one RPT does not change the other.
•
The value of PA is always in Pascal [Pa]. The value of PA
must be entered in the pressure unit Pascal [Pa]. If your
calibration data is in a unit other than Pascal, the calculated
value of PA will be in the other unit and must be converted to
Pascal before it is entered (see Section 7.1.1 for pressure
unit conversions). The value of PM is dimensionless.
5.2.5
RPT Calibration/Adjustment Procedure Without Using
CalTool for RPTs Software
Caution
Most users should use CalTool for RPTs software to assist in the
calibration of reference pressure transducers. CalTool for RPTs
and its documentation are provided on the General Accessories Disk
delivered with . Whether using CalTool or not, before
proceeding to calibrate RPTs, Sections 5.2.1 through 5.2.4 of
this manual should be reviewed thoroughly.