™/S, MOLBLOC TERMINAL OPERATION AND MAINTENANCE MANUAL
© 1995 - 2010 Fluke Corporation, DH Instruments Division
Page 4
Gases Supported
molbloc-L and molbloc-S
Nitrogen (N2), dry Air, humid Air (molbloc-S only) , Argon (Ar), Butane
(Butn), Carbon Monoxide (CO), Helium (He), Oxygen (O2), Carbon
Dioxide (CO2), Carbon Tetrafluoride (CF4), Octofluorocyclobutane
(C4F8), Ethane (C2H6), Ethylene (C2H4), Fluoroform (CHF3),
Hexafluoroethane (C2F6), Hydrogen (H2), Methane (CH4), Nitrous
Oxide (N2O), Propane (C3H8), Sulfur Hexafluoride (SF6), Xenon (Xe)
Flow Ranges
See Section 1.2.5.
Valve Driver Option
(8) 12 V outputs.
Each output can sink,500 mA at 12 V, max. 1 A total (see Section 7.2).
CE Conformance
Available. Must be specified.
1.2.2
Reference Pressure Transducer (RPT) Specifications
Type
Oscillating quartz crystal with mechanical bellows
Calibrated Range
A350K
20 to 300 kPa absolute (2.9 to 44 psia)
A700K
20 to 600 kPa absolute (2.9 to 87 psia)
A1.4M
100 to 1200 kPa absolute (14.5 to 174 psia)
A2M
100 to 2000 kPa absolute (14.5 to 290 psia)
Resolution
A350K
0.4 Pa (0.00005 psi)
A700K
0.7 Pa (0.0001 psi)
A1.4M
1.4 Pa (0.0002 psi)
A2M
2 Pa (0.0003 psi)
Accuracy
A350K
Absolute pressure
1
± (0.01 % of reading or 0.003 % Q-RPT span,
whichever is greater)
Differential
± (4.2 Pa (0.0006 psi) or 0.026 % ΔP,
whichever is greater)
A700K
Absolute pressure
1
± (0.01 % of reading or 0.003 % Q-RPT span,
whichever is greater)
Differential
± (8.4 Pa (0.0012 psi) or 0.032 % ΔP,
whichever is greater)
A1.4M
Absolute pressure
1
± (0.01 % of reading or 0.003 % Q-RPT span,
whichever is greater)
A2M
Absolute pressure
1
± (0.01 % of reading or 0.003 % Q-RPT span,
whichever is greater)
1
With regular use of Autozero. Add 0.005 % of Q-RPT span for one
year without use of AutoZero,
1.2.3
Temperature Measurement Specifications
Specifications are for molbloc mounted Platinum Resistance Thermometers (PRT) combined
with resistance measurement system and temperature calculation.
Range
0 to 40 °C
Accuracy
± 0.05 °C
Resolution
0.01 °C
The internal resistance measurement system is automatically calibrated using
reference 100 and 110
Ω
(± 0.01 %) resistors (see Section 5.4).