M O L B L O C / M O L B O X ™ L O W M A S S F L O W C A L I B R A T I O N S Y S T E M A F M E T C A L
C O N F I G U R A T I O N S
I N S T A L L A T I O N A N D O P E R A T I O N M A N U A L S U P P L E M E N T
Page
15
©2000-2008
DH Instruments, a Fluke Company
gases
All molblocs are calibrated for use in nitrogen. To make traceable measurements with the same uncertainty in another gas, molblocs
should be calibrated in that gas also. Calibration data for several gases can be stored in each molbloc. All AFMETCAL configuration
system molblocs
except 10 sccm (1E1) range molblocs
are calibrated for air in addition to nitrogen. Air calibrations for 10 sccm
(1E1) ranges are not currently available.
Selecting the active gas from the molbox1 front panel or using COMPASS software automatically activates the molbloc calibration
data for that gas. No special procedure is required other than being sure that the molbox1 gas selection is the gas that is actually flowing.
System accuracy specifications and the molbloc usable ranges listed in Table 3 are not valid for gases for which
molblocs have not been calibrated.
MOLBLOC supply pressure
molblocs with high pressure calibrations should always be used with an inlet pressure of 76 psia (525 kPaa). AFMETCAL
configuration molblocs are actually calibrated to operate with a range of inlet pressures from 47 to 76 psia (325 to 525 kPaa) but since
the molbloc verification is performed at 525 kPaa, best results will be obtained when using the molbloc at this inlet pressure. Keeping
the inlet pressure at this level also gives the user more freedom to set elevated back pressures on the DUT.
The molstic pressure regulator sets the molbloc inlet pressure. (This is not the supply pressure delivered to the molstic inlet. A
higher pressure should be supplied to the molstic inlet according to the molstic Installation Instructions.) The molbloc inlet pressure
can be observed on the molbox front panel or COMPASS software to be sure that the molstic regulator is set correctly. See Section 3.2 for
instructions on setting molbloc inlet pressure.
Do not set the molstic pressure regulator to deliver a pressure greater than 80 psia (600 kPaa) to the molbloc.