Fluke i430-Flexi-TF-II Скачать руководство пользователя страница 5

9. 

Observe the safety instructions when you connect and use the 
i430-Flexi-TF-II. 

10.  Wear protective equipment or de-energize the circuit before you 

disconnect the flexible probe. 

W

 

Warning 

To prevent false and misleading readings on the instrument, 
the setup for the current probe in the 430-series Power 
Quality Analyzer must be the i430 Flex (on Fluke 430 series I) 
or i430-Flexi-TF-II (on Fluke 430 series II). 

Maintenance 

Before each use to assure continued safety, inspect the Probe and its 
latching system for any damage. Pay particular attention to the insulation 
surrounding the Probe. A Probe under warranty will be promptly repaired 
or replaced (at Fluke's discretion) and returned at no charge. 

XW

 

Warning 

To prevent possible electrical shock, fire, or personal 
injury: 

 

Remove the input signals before you clean the 
Product. 

 

Have an approved technician repair the Product. 

The flexible probe and its latching system require no special care. 

For routine maintenance: 

1. 

Ensure that no foreign body obstructs the latching mechanism. 

2. 

Clean the Probe with a damp cloth and a mild detergent. Do not use 
abrasives, solvents, or alcohol. 

3. 

Do not immerse the Probe in liquids. 

If the i430-Flexi-TF-II does not work 

If the Probe does not perform properly, use the following steps to help 
isolate the problem: 

1. 

Inspect the coupling system for any damage. 

If any foreign material is present, the coupling system will not close 
properly and errors will result. 

2. 

Inspect the cable between the flexible probe and the Power Quality 
Analyzer for any damage. 

3. 

Check if the i430-Flexi-TF-II is the selected current probe on the 
Power Quality Analyzer (Clamp i430-Flexi-TF-II). 

4. 

Verify that the function and range selection on the Power Quality 
Analyzer is correct. 

Содержание i430-Flexi-TF-II

Страница 1: ...heck the user documentation of your test tool for compatibility before you use this current sensor Available versions are i430 Flexi TF II 61 cm 24 inch i430 Flexi TFII 48 122 cm 48 inch Contacting Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 36...

Страница 2: ...ductors are exposed Limit operation to the specified measurement category voltage or current ratings De energize the circuit or wear personal protective equipment in compliance with local requirements before you apply or remove the Product Do not touch voltages 30 V ac 42 V ac peak or 60 V dc Do not use the Product around explosive gas vapor or in damp or wet environments Do not use the Product if...

Страница 3: ...cuits connected directly to utilization points socket outlets and similar points of the low voltage MAINS installation Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Category IV is applicable to test and measuring circuits connected at the source of the building s low voltage MAIN...

Страница 4: ...low these instructions 1 Connect the BNC connector of the i430 Flexi TF II to the input on the Fluke 430 series Power Quality Analyzer 2 Wear protective equipment or de energize the circuit before you connect the flexible probe around the conductor See Figure 1 C B A 1 3 1 3 1 3 D hcf157 eps Figure 1 3 Re energize the circuit if not energized 4 Center the conductor perpendicularly inside the flexi...

Страница 5: ...y Remove the input signals before you clean the Product Have an approved technician repair the Product The flexible probe and its latching system require no special care For routine maintenance 1 Ensure that no foreign body obstructs the latching mechanism 2 Clean the Probe with a damp cloth and a mild detergent Do not use abrasives solvents or alcohol 3 Do not immerse the Probe in liquids If the ...

Страница 6: ...f reading at 10 to 100 of range Noise 1 mV rms at 10 Hz to 10 kHz Additional errors temperature coefficient over operating temperature range for 24 inch model 0 02 of reading C 0 011 F for 48 inch model 0 035 of reading C 0 02 F with position of conductor in the probe window see Figure 1 Zone A Zone B Zone C 0 5 to 30 A 0 5 of reading 1 0 of reading 2 0 of reading 1 count or 0 1 A whichever is gre...

Страница 7: ...er fuses disposable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To obtain service during the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Cen...

Отзывы: