i400s 

Calibration Information 

Safety Information 

 

XW

 Read First: Safety Information 

To ensure safe operation and service of the Current Clamp, follow these 

instructions: 

 

Read the operating instructions before use and follow all safety 

instructions. 

 

Use the Current Clamp only as specified in the operating instructions, 

otherwise the clamp’s safety features may not protect you. 

 

Adhere to local and national safety codes. Individual protective equipment 

must be used to prevent shock and arc blast injury where hazardous live 

conductors are exposed. 

 

Do not hold the Current Clamp anywhere beyond the tactile barrier, see 

Figure 1. 

 

Before each use, inspect the Current Clamp. Look for cracks or missing 

portions of the clamp housing or output cable insulation. Also look for 

loose or weakened components. Pay particular attention to the insulation 

surrounding the jaws.  

 

Never use the Current Clamp on a circuit with voltages higher than 1000 V 

CAT III or 600 V CAT IV.  

 

CAT III equipment is designed to protect against transients in 

equipment in fixed-equipment installations, such as distribution 

panels, feeders and short branch circuits, and lighting systems in 

large buildings. 

 

CAT IV equipment is designed to protect against transients from the 

primary supply level, such as an electricity meter or an overhead or 

underground utility service. 

 

Use extreme caution when working around bare conductors or bus bars. 

Contact with the conductor could result in electric shock. 

 

Use caution when working with voltages above 60 V dc or 30 V ac. Such 

voltages pose a shock hazard. 

Electrical Symbols 

 

The symbols in Table 1 appear in this document and on the Current Clamp. 

Table 1. Electrical Symbols 

Symbol 

Explanation 

,

 

Application around and removal from hazardous live conductors is permitted. 

T

 

Product is protected by double insulation. 

W

 

Risk of Danger.  Important information. See Instruction Sheet. 

X

 

Hazardous voltage. 

)

 

Conforms to relevant Canadian Standards Association directives. 

P

 

Conforms to relevant European Union directives. 

www.

.com

[email protected]

1.800.561.8187

Содержание i400s

Страница 1: ...se of qualified personnel only The i400s Calibration Information provides the information necessary to verify the performance and adjust the calibration of the Fluke i400s AC Current Clamp hereafter known as the Current Clamp The following information is included in this document Safety Information and Electrical Symbols Specifications Maintenance Performance Tests Calibration Adjustment User Repl...

Страница 2: ...an 1000 V CAT III or 600 V CAT IV CAT III equipment is designed to protect against transients in equipment in fixed equipment installations such as distribution panels feeders and short branch circuits and lighting systems in large buildings CAT IV equipment is designed to protect against transients from the primary supply level such as an electricity meter or an overhead or underground utility se...

Страница 3: ...23 5 C 20 to 75 RH conductor centered in jaw opening no DC component no adjacent conductor 40 A Range 400 A Range Measurement Range 0 5 A to 40 A 5 A to 400 A Output 10 mV A 1 mV A Accuracy 45 Hz to 400 Hz 2 0 015 A 2 0 04 A Phase Shift 45 Hz to 400 Hz 0 5 A to 1 A 1 A to 5 A 5 A to 10 A 10 A to 20 A 20 A to 40 A 40 A to 400 A Unspecified 4 3 3 2 NA NA NA Unspecified 2 2 1 5 Crest Factor 3 3 to 30...

Страница 4: ...WxH 150 x 70 x 30 mm 5 09 x 2 75 x 1 18 in Weight 114 g 4 oz Maintenance XW Warning To avoid possible electric shock or personal injury Before each use inspect the Current Clamp Look for cracks or missing portions of the clamp housing and output cable insulating cover and for loose or weakened components Pay particular attention to the insulation surrounding the clamp jaws Do not use a damaged Cur...

Страница 5: ...unction against the Clamp s specifications If the Current Clamp fails any part of the test calibration adjustment and or repair is indicated In the performance tests the Current Clamp is referred to as the unit under test UUT Required test equipment is defined in Table 2 Table 2 Required Test Equipment Equipment Required Characteristics Recommended Model Calibrator AC Current 0 8 A Max Accuracy 0 ...

Страница 6: ...00 Hz 0 4 A 400 Hz 19 56 20 44 1 mV A 400 A 50 Hz 8 0 A 50 Hz 391 96 408 04 1 mV A 400 A 100 Hz 8 0 A 100 Hz 391 96 408 04 10 mV A 2 A 50 Hz 40 mA 50 Hz 19 45 20 55 10 mV A 2 A 400 Hz 40 mA 400 Hz 19 45 20 55 10 mV A 40 A 50 Hz 0 8 A 50 Hz 391 85 408 15 10 mV A 40 A 100 Hz 0 8 A 100 Hz 391 85 408 15 bbc01f eps Figure 2 Current Accuracy Test Connections www com information itm com 1 800 561 8187 ...

Страница 7: ...t point VR1 3 Using a small flat head screwdriver adjust VR1 for a meter reading of 99 9 mV to 100 1 mV VR1 bbc02f eps Figure 3 i400s Calibration Point User Replaceable Parts The Current Clamp contains no user replaceable parts The following user documentation is available for this product Description Part No Qty i400s Instruction Sheet 2282671 1 i400s Calibration Information this document 2414327...

Страница 8: ...the warranty period To obtain warranty service contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no risk for damage in transit Following warranty repair the product will be returned to Buyer transportation prepai...