i310s English Instruction Sheet 

Page  

Current Measurement 

1.

 

Switch on the Clamp to the required current range and 
check that the LED is lit. 

2.

 

Connect the output lead to an oscilloscope, multimeter, 
or other measuring equipment. 

3.

 

If necessary, adjust the Clamp output voltage to zero as 
described in section Zero Adjustment

4.

 

Clamp the jaw around the conductor ensuring a good 
contact between the closing faces of the jaws. 

5.

 

Observe and take measurements as required. Positive 
output indicates that the current flow is in the direction 
shown by the arrow on the Clamp. 

Maintenance 

Cleaning 

Clean the case periodically by wiping it with a damp cloth and 
detergent. Do not use abrasive cleaners or solvents. Do not 
immerse the Clamp in liquids. 

Battery Replacement 

WX

 Warning 

To avoid personal injury, always ensure the Clamp 
is removed from any live electric circuit, and leads 
are disconnected before removing the battery cover. 

Never operate the Clamp without the battery cover 
fitted. 

The red LED will flash when the minimum operating voltage is 
approached. Refer to Figure 1. Use the following procedure: 

1.

 

Unclamp from the conductor, turn it off using the On – 
Off switch and disconnect the output leads, from 
external equipment. 

2.

 

Loosen the captive screw that secures the battery 
cover. Lift the cover through 30° and pull it clear of the 
Clamp body as shown in Figure 1. The battery is then 
accessible. Replace the battery and re-fit the battery 
cover and fasten the screw. 

Note 

Replacement with other than the specified type of 
battery will invalidate the warranty. 

Fit only the type 9 V PP3 Alkaline (MN 1604). 

 

Содержание i310s

Страница 1: ...l tool for use in inverters switch mode power supplies industrial controllers automotive diagnostics and other applications requiring current measurements and or waveform analysis Symbols The table below lists the symbols used on the Clamp and or in this manual Symbol Description Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information W Important ...

Страница 2: ...is removed from any live electric circuit and leads are disconnected before removing the battery cover Use the Clamp only as specified in the operating instructions otherwise the safety features may not protect you Adhere to local and national safety codes Individual protective equipment must be used to prevent the shock and arc blast injury where hazardous live conductors are exposed Do not hold ...

Страница 3: ...0 A and 300 A ac rms or 45 A and 450 A dc Inrush Current 600 A ac rms Max Output Sensitivity 10 mV A 30 A 1 mV A 300 A Accuracy 30 A range 1 of reading 50 mA 300 A range 1 of reading 300 mA 25 C Bandwidth dc to 1 kHz Bandwidth to Meet Accuracy Specification 1 kHz Phase Shift below 1 kHz 2 degrees Resolution 50 mA 30 A 100 mA 300 A Load impedance 10 k Ω and 100 pF Conductor Position Sensitivity 1 5...

Страница 4: ...2 meters Operating Temperature Range 10 to 50 C 14 to 122 F Storage Temperature Range with battery removed 20 to 85 C 4 to 185 F Operating Humidity 15 to 85 non condensing Weight 250 g 8 812 oz Safety Standards EN 61010 1 2001 EN 61010 2 032 2002 EN 61010 031 2002 300 V rms Category III Pollution Degree 2 Use of the Clamp on non insulated conductors is limited to 300 V ac rms or dc and frequencies...

Страница 5: ...ange Typical Frequency Response i310s evy01_4 eps Typical Frequency Response Amps 2 1 5 1 0 5 0 0 5 1 1 5 2 Error of Reading Typical Accuracy i310s 30A range evy02 eps Typical Frequency Response Amps 2 1 5 1 0 5 0 0 5 1 1 5 2 Error of Reading Typical Accuracy i310s 300A range evy03 eps Typical Accuracy Curve ...

Страница 6: ...nector s appears to be damaged or if a malfunction of the instrument is suspected Switch On Switch the Clamp to the required current range and check that the LED is lit The LED starts flashing when the battery voltage is too low for normal operation and warns the user that it requires changing This procedure is described below Zero Adjustment The output zero offset voltage of the Clamp may change ...

Страница 7: ...ents Do not immerse the Clamp in liquids Battery Replacement WX Warning To avoid personal injury always ensure the Clamp is removed from any live electric circuit and leads are disconnected before removing the battery cover Never operate the Clamp without the battery cover fitted The red LED will flash when the minimum operating voltage is approached Refer to Figure 1 Use the following procedure 1...

Страница 8: ...ke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since...

Отзывы: