i310s English Instruction Sheet 

Page  

Safety Instructions 

Please read this section carefully. It will make you familiar with 
the most important safety instructions for handling your product. 
In this instruction sheet, a Warning identifies conditions and 
actions that pose hazard(s) to the user. A Caution identifies 
conditions and actions that may damage the calibrator or the test 
instruments. 

WX

 

Warning 

The Clamp may only be used and handled by 
qualified personnel. To avoid personal injury, follow 
these precautions: 

 

To avoid electric shock, use caution during 
installation and use of this product; high 
voltages and currents may be present in circuit 
under test. 

 

Do not use the Clamp if damaged. Always 
connect to display device before it is installed 
around the conductor. 

 

Always ensure the Clamp is removed from any 
live electric circuit, and leads are disconnected 
before removing the battery cover. 

 

Use the Clamp only as specified in the 
operating instructions; otherwise the safety 
features may not protect you. 

 

Adhere to local and national safety codes. 
Individual protective equipment must be used 
to prevent the shock and arc blast injury where 
hazardous live conductors are exposed. 

 

Do not hold the Clamp anywhere beyond the 
tactile barrier. 

 

Before each use, inspect the Clamp. Look for 
cracks or missing portions of the housing or 
output cable insulation. Also look for loose or 
weakened components. Pay particular attention 
to the insulation surrounding the jaws.  

 

Use caution when working with voltages above 
60 V dc, 30 V ac rms or 42 V ac peak. Such 
voltages pose a shock hazard.  

 

Use of this equipment is designed to protect 
against transients in equipment in fixed 
equipment installations, such as distribution 
panels, feeders and short branch circuits, and 
lighting systems in large buildings. 

Содержание i310s

Страница 1: ...l tool for use in inverters switch mode power supplies industrial controllers automotive diagnostics and other applications requiring current measurements and or waveform analysis Symbols The table below lists the symbols used on the Clamp and or in this manual Symbol Description Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information W Important ...

Страница 2: ...is removed from any live electric circuit and leads are disconnected before removing the battery cover Use the Clamp only as specified in the operating instructions otherwise the safety features may not protect you Adhere to local and national safety codes Individual protective equipment must be used to prevent the shock and arc blast injury where hazardous live conductors are exposed Do not hold ...

Страница 3: ...0 A and 300 A ac rms or 45 A and 450 A dc Inrush Current 600 A ac rms Max Output Sensitivity 10 mV A 30 A 1 mV A 300 A Accuracy 30 A range 1 of reading 50 mA 300 A range 1 of reading 300 mA 25 C Bandwidth dc to 1 kHz Bandwidth to Meet Accuracy Specification 1 kHz Phase Shift below 1 kHz 2 degrees Resolution 50 mA 30 A 100 mA 300 A Load impedance 10 k Ω and 100 pF Conductor Position Sensitivity 1 5...

Страница 4: ...2 meters Operating Temperature Range 10 to 50 C 14 to 122 F Storage Temperature Range with battery removed 20 to 85 C 4 to 185 F Operating Humidity 15 to 85 non condensing Weight 250 g 8 812 oz Safety Standards EN 61010 1 2001 EN 61010 2 032 2002 EN 61010 031 2002 300 V rms Category III Pollution Degree 2 Use of the Clamp on non insulated conductors is limited to 300 V ac rms or dc and frequencies...

Страница 5: ...ange Typical Frequency Response i310s evy01_4 eps Typical Frequency Response Amps 2 1 5 1 0 5 0 0 5 1 1 5 2 Error of Reading Typical Accuracy i310s 30A range evy02 eps Typical Frequency Response Amps 2 1 5 1 0 5 0 0 5 1 1 5 2 Error of Reading Typical Accuracy i310s 300A range evy03 eps Typical Accuracy Curve ...

Страница 6: ...nector s appears to be damaged or if a malfunction of the instrument is suspected Switch On Switch the Clamp to the required current range and check that the LED is lit The LED starts flashing when the battery voltage is too low for normal operation and warns the user that it requires changing This procedure is described below Zero Adjustment The output zero offset voltage of the Clamp may change ...

Страница 7: ...ents Do not immerse the Clamp in liquids Battery Replacement WX Warning To avoid personal injury always ensure the Clamp is removed from any live electric circuit and leads are disconnected before removing the battery cover Never operate the Clamp without the battery cover fitted The red LED will flash when the minimum operating voltage is approached Refer to Figure 1 Use the following procedure 1...

Страница 8: ...ke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since...

Отзывы: