Specifications 

Accuracy is specified as 

±

 ([% of reading] + [number of least significant digits] ). All ranges are autoranging. Accuracy is 

specified from 5 % to 100 % of  the range obtained by autoranging, from 18 

°

C to 28 

°

C. 

AC Voltage 

Range  

Resolution 

Accuracy 

[1][2][3]

 

45 Hz to 500 Hz 

500 Hz to 1 kHz 

6.000 V 

0.001 V 

 

1.0 % + 3 

 

2.0 % + 3 

60.00 V 

0.01 V 

600.0 V 

0.1 V 

1000 V 

1 V 

[1] 

Crest factor of 

3 at full scale to a maximum of 4000 counts, decreasing linearly to crest factor 

1.5 at full scale. 

[2] 

For non-sinusoidal waveforms, add -(2 % of r 2 % full scale)  typical, for crest factor to a maximum of 3. 

[3]     Do not exceed 10

6

 

V-Hz 

General 

LCD w/Backlight .................................................... 3 ½ digits, 6000 counts, 4/sec update rate 
Log Rate/Interval ................................................... adjustable by PC from 1 second to 1 hour, default is 1 minute 
Memory ................................................................. Record up to 65,000 readings 
Battery Type .......................................................... 2 AA, NEDA 15 A, IEC LR6 
Battery Life ............................................................ 400 hours 
RF Communications ............................................. 2.4 GHz ISM Band 
RF Communication Range.................................... 20 m (65.61 ft) 

Operating Temperature ......................................... -10 

°

C to +50 

°

C, (14 

°

F to 122 

°

F) 

Storage Temperature ............................................ -40 

°

C to +60 

°

C, (-40 

°

F to 140 

°

F) with batteries removed 

Operating Humidity ............................................... 90 % from 0 

°

C to 35 

°

C, 45 % to 40 

°

C, 45 % to 50 

°

 

(90 % from 32 

°

F to 95 

°

F, 45 % to 104 

°

F, 45 % to 122 

°

F) 

Operating Altitude…………………………………….2,000 m (6561.68 ft) 
Storage Altitude………………………………………12,000 m (39370.079 ft) 
Ingress Protection (IP) rating ................................ IP42 
Radio Frequency Certification............................... FCC Part 15 Subpart C Sections 15.207, 15.209, 

15.249, FCCID : T68-FWCS IC:6627A-FWCS 

Safety .................................................................... IEC 61010-1: 600V CATIV/1000V CATIII, Pollution Degree 2 
EMC 

Applies to use in Korea only ........................... Class A Equipment (Industrial Broadcasting & Communication 

Equipment) [1] 

 

[1] This product meets requirements for industrial (Class A) 

 

electromagnetic wave equipment and the seller or user should take 

 

notice of it. This equipment is intended for use in business 

 

environments and is not to be used in homes. 

EM Environment ................................................... IEC 61236-1: Portable 

Temperature Coefficient ....................................... 0.1 X (specified accuracy)/ 

°

C (<18 

°

C or >28 

°

C) 

Size ....................................................................... (160 tall x 66 wide x 38 deep) mm

 

 

(6.3 tall x 2.6 wide x 1.5 deep) in 

Weight ...................................................................  0.255 kg (9 oz)

Содержание CNX v3000

Страница 1: ...012 2012 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies CNX v3000 Wireless AC Voltage Calibrat...

Страница 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Страница 3: ...equired Equipment 5 Performance Tests 5 Test the Display 5 Backlight 6 Keypad Test 6 AC Voltage Test 6 Before Calibration Adjustment 7 Maintenance Mode 7 Password Entry 8 Change the Password 8 Restore...

Страница 4: ...CNX v3000 Calibration Manual ii...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Required Equipment 5 3 Performance Tests 6 4 Calibration Adjustment 10 5 User Replaceable Parts 12...

Страница 6: ...CNX v3000 Calibration Manual iv...

Страница 7: ...v List of Figures Figure Title Page 1 All Segments of the Display 5 2 Performance Test Connections 7 3 Calibration Password Reset 9 4 Battery Replacement 12...

Страница 8: ...CNX v3000 Calibration Manual vi...

Страница 9: ...luke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31...

Страница 10: ...urements Have an approved technician repair the Product Use only specified replacement parts Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product pro...

Страница 11: ...o test and measuring circuits connected directly to utilization points socket outlets and similar points of the low voltage MAINS installation CAT III Measurement Category III is applicable to test an...

Страница 12: ...4 GHz ISM Band RF Communication Range 20 m 65 61 ft Operating Temperature 10 C to 50 C 14 F to 122 F Storage Temperature 40 C to 60 C 40 F to 140 F with batteries removed Operating Humidity 90 from 0...

Страница 13: ...test procedures unless the Product is fully assembled The performance tests verify the full operation of the Product and measure the accuracy of each function against Product specifications If the Pro...

Страница 14: ...up the Calibrator as necessary Refer to its specifications 4 Let the temperature of the unit under test UUT become stable to room temperature To do the ac voltage test 1 Connect the Product to Calibra...

Страница 15: ...Adjustment Before the Product calibration can be adjusted you must go through the maintenance mode menu and enter your password Maintenance Mode The Product maintenance mode can be used to set differ...

Страница 16: ...correct password is entered C 01 is shown If the incorrect password is entered is shown and the password must be correctly entered Change the Password Note If you change the password and then lose it...

Страница 17: ...r handling them 3 Using the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HAND...

Страница 18: ...E EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE...

Страница 19: ...Phillips screwdriver remove the bottom case screws Two of the screws are inside of the battery door 3 Keep the pca in the top case 4 Apply 3 0 V across the battery contacts on the pca Note the polarit...

Страница 20: ...on After confirmation the Product goes to the subsequent calibration step Note After you push wait until the calibration step number advances before you change the Calibrator source Some adjustment st...

Страница 21: ...uct or to equipment under test Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period Be sure that the battery polarity is correct to prevent ba...

Страница 22: ...Part Number Description Qty 4130305 FLK 3000 2003 DOOR BATTERY 1 1881997 TPAK Meter Hanging Kit 1 2061552 TL224 4201 175 263 011 TEST LEADS RA2S 2 1610159 AC285 5001 01 175 276 011 AC285 LARGE ALLIGA...

Отзывы: