Wireless AC Voltage
Performance Tests
5
Required Equipment
The equipment in Table 2 is necessary for performance tests and calibration adjustment.
Table 2. Required Equipment
Recommended Equipment
Measurement Function
Accuracy
Fluke 5500A Calibrator
(or equivalent)
AC Volts
6 mV to 600 V
±
0.25 % @ 45 Hz to 1 kHz
Performance Tests
Warning
To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the full operation of the Product and measure the accuracy
of each function against Product specifications. If the Product fails a part of the test,
calibration adjustment and/or repair is necessary. See “Calibration Adjustment”.
Test the Display
To verify that all segments of the display function:
1.
With the Product off, push and hold
.
2.
Push
while you keep
pushed until all of the display segments are shown. See
Figure 1.
hby002.eps
Figure 1. All Segments of the Display
If segments of the display are missing, repair is necessary. See “Contact Fluke”.