Index 2MF
Users Manual
7-6
esl063.eps
3.
To raise the Pulse Amp percentage (AC factors) in the simulated blood condition,
press the softkey
+
:
esl064.eps
The amplitude will
increase
in increments of 1% with every softkey press.
4.
To lower the percentage of AC factors in the simulated blood condition, press the
–
softkey:
esl065.eps
The amplitude will
decrease
by 1% with every softkey press down to 1 %. From 1 %
to 0.1 % the value will decrease by 0.1 %. Below 0.1 %, the value will decrease by
0.025 % per button press. Inputs below 6 % are rounded internally to the nearest
0.024 %, and above 6 % are rounded to the nearest 0.072 %. The displayed value is
the rounded value nearest the value displayed per the preceding increment/decrement
rules. If the pulse amplitude rate exceeds the oximeter alarm rate, the alarm will
sound on the pulse oximeter being tested after a period of time.
5.
To print the results of the alarm, press the
PRT
softkey.
esl067.eps
The results will be sent to the RS232 or Centronics port.
Содержание Biomedical Index 2MF
Страница 8: ...Index 2MF Users Guide iv...
Страница 10: ...Index 2MF Users Guide vi...
Страница 12: ...Index 2MF Users Guide viii...
Страница 18: ...Index 2MF Users Guide 1 6...
Страница 28: ...Index 2MF Users Guide 4 4...
Страница 34: ...Index 2MF Users Manual 5 6...
Страница 72: ...Index 2MF Users Guide 9 6...
Страница 76: ...Index 2MF Users Guide 10 4...
Страница 86: ...Index 2MF Users Guide 12 4...
Страница 88: ...Index 2MF Users Guide...
Страница 94: ...Index 2MF Users Guide A 6...
Страница 98: ...Index 2MF Users Guide C 2...
Страница 112: ...Index 2MF Users Guide F 4...
Страница 126: ...Index 2MF Users Guide H 6...