Index 2MF
Users Guide
D-2
Dynamic Test
Test Format........................................................... Photo detector/diode response to both the red and infrared light
generated by the probe when pulsed by an internal test signal.
Test Signal.......................................................... Pulsed between the two LEDs, constant current level at 1.0 mA
Test Results........................................................ Nominal range of 0 to 20,000
Checksum
.............................................................. The sum of all locations in the program chip. For service only.
Note
The above specifications are for pre-programmed makes only, 20 % pulse
amplitude SpO2. Results for Presets, custom makes, and various pulse
amplitudes are unspecified and depend on the algorithm of the specific
pulse oximeter being tested.
Содержание Biomedical Index 2MF
Страница 8: ...Index 2MF Users Guide iv...
Страница 10: ...Index 2MF Users Guide vi...
Страница 12: ...Index 2MF Users Guide viii...
Страница 18: ...Index 2MF Users Guide 1 6...
Страница 28: ...Index 2MF Users Guide 4 4...
Страница 34: ...Index 2MF Users Manual 5 6...
Страница 72: ...Index 2MF Users Guide 9 6...
Страница 76: ...Index 2MF Users Guide 10 4...
Страница 86: ...Index 2MF Users Guide 12 4...
Страница 88: ...Index 2MF Users Guide...
Страница 94: ...Index 2MF Users Guide A 6...
Страница 98: ...Index 2MF Users Guide C 2...
Страница 112: ...Index 2MF Users Guide F 4...
Страница 126: ...Index 2MF Users Guide H 6...