Troubleshooting

If problems arise while operating the 9135, this section provides some sugges-
tions that may help you solve the problem. If, after considering the following
suggestions, the problem still cannot be resolved, call an Authorized Service
Center for assistance. Opening the unit without contacting an Authorized Ser-
vice Center may void the warranty.

Problem

Causes and Solutions

Incorrect temperature reading

Check that the green ready LED is lit showing that the IR target has
stabilized to the set-point.
Check that the IR thermometer is close enough to the IR target.
Check that the IR thermometer is set to 0.95 emmisivity.
Check that the IR target surface is clean and free of particles.
Check the temperature with a calibrated 3.18 mm. (0.125") diameter
RTD to verify the temperature of the IR target.
Have the instrument calibrated regularly.

None of the LEDs light

Check that the power cord is plugged in and the power switch is on.
Check that the mains supply is the correct voltage as indicated on the
back of the calibrator.
The internal fuse may be blown or the calibrator may be damaged in
which case the instrument should be returned. Contact a Hart Scien-
tific Authorized Service Center for assistance.

The ready indicator stays dark

Even under normal circumstances it can take up to ten minutes to
heat up to and stabilize at a higher temperature and up to 40 minutes
to cool down to and stabilize at a lower temperature.
Check that the mains supply is the correct voltage as indicated on the
back of the calibrator.
Make sure the mains supply is reasonably free from surges,
drop-outs, and interference.

11.1

CE Comments

11.1.1

EMC Directive

Hart Scientific’s equipment has been tested to meet the European Electromag-
netic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration
of Conformity for your instrument lists the specific standards to which the unit
was tested.

11.1.2

Low Voltage Directive (Safety)

In order to comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN
61010-1) and IEC 1010-2-010 (EN 61010-2-010) standards.

27

11 Troubleshooting

CE Comments

Содержание 9135

Страница 1: ...9135 Infrared Thermometer Calibrator User s Guide Rev 5A2604 Hart Scientific...

Страница 2: ...entific Division 799 E Utah Valley Drive American Fork UT 84003 9775 USA Phone 1 801 763 1600 Telefax 1 801 763 1010 E mail support hartscientific com www hartscientific com Subject to change without...

Страница 3: ...cations 9 3 2 Environmental Conditions 9 3 3 Warranty 10 4 Quick Start 11 4 1 Unpacking 11 4 2 Set up 11 4 3 Power 11 5 Parts and Controls 13 5 1 Back Panel 13 5 2 Bottom 13 5 3 Top Panel 13 5 4 Right...

Страница 4: ...9 Calibration Procedure 23 10 Maintenance 25 11 Troubleshooting 27 11 1 CE Comments 27 11 1 1 EMC Directive 27 11 1 2 Low Voltage Directive Safety 27 ii...

Страница 5: ...he instrument or in this manual Symbol Description AC Alternating Current AC DC Battery CE Complies with European Union Directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface Burn Ha...

Страница 6: ...to the user Caution identifies conditions and actions that may damage the instru ment being used 1 2 1 Warnings To avoid personal injury follow these guidelines DO NOT operate this unit without a pro...

Страница 7: ...l to start a fire Operate the instrument in room temperatures between 5 40 C 41 104 F Allow sufficient air circulation by leaving at least three inches of space between the instrument and nearby objec...

Страница 8: ...ail ServiceDesk fluke nl Fluke Int l Corporation Service Center Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004 PRC CHINA Phone 86 10 6 512 3436 Telefax 86...

Страница 9: ...singa fluke com When contacting these Service Centers for support please have the following information available Model Number Serial Number Voltage Complete description of the problem 5 1 Before You...

Страница 10: ...roportional integral tempera ture controller maintains the temperature of the plate The set point can be con veniently selected by the one button control to 50 C 100 C or 150 C One of three yellow LED...

Страница 11: ...on Degree 2 per IEC1010 1 3 2 Environmental Conditions Although the instrument has been designed for optimum durability and trou ble free operation it must be handled with care The instrument should n...

Страница 12: ...g To exercise this warranty the purchaser must forward the product after calling or writing a Hart Scientific Authorized Service Center see Section 1 3 for au thorization The Service Centers assume NO...

Страница 13: ...ck the power connection 4 3 Power The calibrator is configured for either 115 VAC 10 50 60 Hz or 230 VAC 10 50 60 Hz using switches underneath There are two switches and both must be set to the correc...

Страница 14: ...he mains supply must be reasonably stable and free from surges dropouts and interference that can affect the temperature stability of the IR target 9135 Infrared Thermometer Calibrator User s Guide 12...

Страница 15: ...1 IR Target The large hole in the top panel gives access to the IR target The IR target is a metal plate that is coated with a special paint with an emmisivity of 0 95 The IR target is controlled at t...

Страница 16: ...e the 9135 against a standard RTD see Section 9 Calibration Access Hole There is a small oblong hole on the right side panel These gives access to two potentiometers used to calibrate the 9135 DO NOT...

Страница 17: ...entiometers unless qualified to do so as part of a regular mainte nance and calibration schedule see Section 9 15 5 Parts and Controls Right Side Panel Prop Calibration Access Hole Test Well Figure 2...

Страница 18: ...erature stabilizes To select the 100 C set point press the select button once and observe the 100 C set point LED light up To select the 150 C set point press the select button once more and observe t...

Страница 19: ...s the temperature of the IR target using a thermistor sensor imbedded in the plate It compares the resistance of the thermistor to preset set point resistances to determine whether the IR target shoul...

Страница 20: ...d When calibrating IR thermometers the thermometer must see only the IR target of the 9135 The thermometer must be pointed directly at the tar get at approximately a 90 angle and with a close enough p...

Страница 21: ...peration of the 9135 and especially affect the ability to calibrate the instrument in the future DO NOT attempt to insert fluid grease or any objects other than clean thermometers into the well Insert...

Страница 22: ...shows the temperature is close to 50 C within 0 2 C Adjust the potentiometer clock wise to increase the temperature or counter clockwise to decrease the temperature 4 Set the set point to 150 C and l...

Страница 23: ...care Avoid knocking bumping or dropping the calibrator Due to the size and portability of the 9135 it is vulnerable to being shocked and dropped as it is carried around It might be inadvertently drag...

Страница 24: ...tor cause loss of accuracy and may cause sticking and damage to your probes In the event that the calibrator should require service or repair please contact a Hart Scientific Authorized Service Center...

Страница 25: ...correct voltage as indicated on the back of the calibrator The internal fuse may be blown or the calibrator may be damaged in which case the instrument should be returned Contact a Hart Scien tific Au...

Отзывы: