background image

9

Calibration Procedure

The 9135 should be calibrated regularly to ensure accuracy. A one-year or less
calibration interval is recommended. Calibration should only be performed by a
qualified and authorized technician.

The calibration of the 9135 entails adjusting the calibration potentiometers in-
side the instrument until the temperature of the IR target is close to the
set-point temperature as observed with a calibrated standard RTD at each of the
three set-points. The following are required:

Standard calibrated RTD or thermistor, 3.18 mm (0.125") diameter and at
least 100 mm (4") sheath length, and a readout device, total uncertainty
0.2°C or better

Potentiometer adjustment tool

The calibration procedure is as follows:

1. Insert the standard RTD or thermistor fully into the test well of the 9135
through the test well access hole in the right panel.

2. Set the set-point to 50°C and let it stabilize. Wait for the green ready LED to
light.

3. Adjust the potentiometer on the right one-half turn at a time, waiting one to
two minutes between adjustments, until the standard RTD or thermistor shows
the temperature is close to 50°C within 0.2°C. Adjust the potentiometer clock-
wise to increase the temperature or counter-clockwise to decrease the
temperature.

4. Set the set-point to 150°C and let it stabilize. Wait for the green ready LED
to light.

5. Adjust the potentiometer on the left one-half turn at a time, waiting one to
two minutes between adjustments, until the standard RTD or thermistor shows
the temperature is close to 150°C within 0.2°C. Adjust the potentiometer clock-
wise to increase the temperature or counter-clockwise to decrease the
temperature.

6. Set the set-point to 100°C and let it stabilize. Wait for the green ready LED
to light.

7. If the standard RTD or thermistor shows the temperature differs from 100°C
by more than 0.5°C but less than 1°C then adjust the potentiometer on the right
one-half turn at a time, waiting one to two minutes between adjustments, until
the standard RTD or thermistor shows the temperature is close to 100°C within
0.25 to 0.5°C. Do not force the temperature to exactly 100°C or the error at the
other two set-points will be excessive. If the temperature differs from 100°C by
more than 1°C then the instrument may require factory service.

8. Check the temperature once more at each of the three set-points.

23

9 Calibration Procedure

Содержание 9135

Страница 1: ...9135 Infrared Thermometer Calibrator User s Guide Rev 5A2604 Hart Scientific...

Страница 2: ...entific Division 799 E Utah Valley Drive American Fork UT 84003 9775 USA Phone 1 801 763 1600 Telefax 1 801 763 1010 E mail support hartscientific com www hartscientific com Subject to change without...

Страница 3: ...cations 9 3 2 Environmental Conditions 9 3 3 Warranty 10 4 Quick Start 11 4 1 Unpacking 11 4 2 Set up 11 4 3 Power 11 5 Parts and Controls 13 5 1 Back Panel 13 5 2 Bottom 13 5 3 Top Panel 13 5 4 Right...

Страница 4: ...9 Calibration Procedure 23 10 Maintenance 25 11 Troubleshooting 27 11 1 CE Comments 27 11 1 1 EMC Directive 27 11 1 2 Low Voltage Directive Safety 27 ii...

Страница 5: ...he instrument or in this manual Symbol Description AC Alternating Current AC DC Battery CE Complies with European Union Directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface Burn Ha...

Страница 6: ...to the user Caution identifies conditions and actions that may damage the instru ment being used 1 2 1 Warnings To avoid personal injury follow these guidelines DO NOT operate this unit without a pro...

Страница 7: ...l to start a fire Operate the instrument in room temperatures between 5 40 C 41 104 F Allow sufficient air circulation by leaving at least three inches of space between the instrument and nearby objec...

Страница 8: ...ail ServiceDesk fluke nl Fluke Int l Corporation Service Center Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004 PRC CHINA Phone 86 10 6 512 3436 Telefax 86...

Страница 9: ...singa fluke com When contacting these Service Centers for support please have the following information available Model Number Serial Number Voltage Complete description of the problem 5 1 Before You...

Страница 10: ...roportional integral tempera ture controller maintains the temperature of the plate The set point can be con veniently selected by the one button control to 50 C 100 C or 150 C One of three yellow LED...

Страница 11: ...on Degree 2 per IEC1010 1 3 2 Environmental Conditions Although the instrument has been designed for optimum durability and trou ble free operation it must be handled with care The instrument should n...

Страница 12: ...g To exercise this warranty the purchaser must forward the product after calling or writing a Hart Scientific Authorized Service Center see Section 1 3 for au thorization The Service Centers assume NO...

Страница 13: ...ck the power connection 4 3 Power The calibrator is configured for either 115 VAC 10 50 60 Hz or 230 VAC 10 50 60 Hz using switches underneath There are two switches and both must be set to the correc...

Страница 14: ...he mains supply must be reasonably stable and free from surges dropouts and interference that can affect the temperature stability of the IR target 9135 Infrared Thermometer Calibrator User s Guide 12...

Страница 15: ...1 IR Target The large hole in the top panel gives access to the IR target The IR target is a metal plate that is coated with a special paint with an emmisivity of 0 95 The IR target is controlled at t...

Страница 16: ...e the 9135 against a standard RTD see Section 9 Calibration Access Hole There is a small oblong hole on the right side panel These gives access to two potentiometers used to calibrate the 9135 DO NOT...

Страница 17: ...entiometers unless qualified to do so as part of a regular mainte nance and calibration schedule see Section 9 15 5 Parts and Controls Right Side Panel Prop Calibration Access Hole Test Well Figure 2...

Страница 18: ...erature stabilizes To select the 100 C set point press the select button once and observe the 100 C set point LED light up To select the 150 C set point press the select button once more and observe t...

Страница 19: ...s the temperature of the IR target using a thermistor sensor imbedded in the plate It compares the resistance of the thermistor to preset set point resistances to determine whether the IR target shoul...

Страница 20: ...d When calibrating IR thermometers the thermometer must see only the IR target of the 9135 The thermometer must be pointed directly at the tar get at approximately a 90 angle and with a close enough p...

Страница 21: ...peration of the 9135 and especially affect the ability to calibrate the instrument in the future DO NOT attempt to insert fluid grease or any objects other than clean thermometers into the well Insert...

Страница 22: ...shows the temperature is close to 50 C within 0 2 C Adjust the potentiometer clock wise to increase the temperature or counter clockwise to decrease the temperature 4 Set the set point to 150 C and l...

Страница 23: ...care Avoid knocking bumping or dropping the calibrator Due to the size and portability of the 9135 it is vulnerable to being shocked and dropped as it is carried around It might be inadvertently drag...

Страница 24: ...tor cause loss of accuracy and may cause sticking and damage to your probes In the event that the calibrator should require service or repair please contact a Hart Scientific Authorized Service Center...

Страница 25: ...correct voltage as indicated on the back of the calibrator The internal fuse may be blown or the calibrator may be damaged in which case the instrument should be returned Contact a Hart Scien tific Au...

Отзывы: