9132 Portable IR Calibrator
User’s Guide
2
Table 3
Serial Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint]
s
set: 999.9 {C or F}
set: 100.00 C
Set current set-point to n
s[etpoint]=n
s=200.0
Instrument
Range
Read temperature
t[emperature]
t
t: 999.9 {C or F}
t: 55.6 C
Read temperature units
u[nits]
u
u: x
u: C
Set temperature units:
u[nits]=c/f
C or F
Set temperature units to Celsius
u[nits]=c
u=c
Set temperature units to Fahrenheit
u[nits]=f
u=f
Read scan mode
sc[an]
sc
scan: {ON or OFF}
scan:ON
Set scan mode
sc[an]=on/off
sc=on
ON or OFF
Read scan rate
sr[ate]
sr
srat: 99.9 {C or
F}/min
srat:12.4C/min
Set scan rate
sr[ate]=n
sr=1.1
.1 to 99.9
Secondary Menu
Read proportional band setting
pr[opband]
pr
pb: 999.9
pb: 15.9
Set proportional band to
n
pr[opband]=n
pr=8.83
Depends on
Configuration
Read heater power
(duty cycle)
po[wer]
po
po: 999.9
po: 1.0
Configuration Menu
Operating Parameters Menu
Read high limit
hl
hl
hl:999
hl:126
Set high limit
hl=n
hl=90
0–126
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
Set serial sampling setting to
n
seconds
sa[mple]=n
sa=0
0 to 999
Set serial duplex mode:
du[plex]=f[ull]/
h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu
Read R0 calibration parameter
r[0]
r
r0: 999.999
r0: 100.578
Set R0 calibration parameter to
n
r[0]=n
r=100.324
90 to 110
Read ALPHA calibration parameter
al[pha]
al
al: 9.9999999
al: 0.0038573
Set ALPHA calibration parameter to
n
al[pha]=n
al=0.0038433
.002 to .005
Read DELTA calibration parameter
de[lta]
de
de:9.99999
de: 1.507