NOTE: As the calibrator temperature increases, the difference between the temperature
sensed by the reference thermometer and the temperature of the face of the target
increases.
2.2
Environmental Conditions
Although the instrument has been designed for optimum durability and trouble-free operation,
it must be handled with care. The instrument should not be operated in an excessively dusty or
dirty environment. Maintenance and cleaning recommendations can be found in the Mainte-
nance section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5-35°C (41-95°F)
ambient relative humidity: 15-50%
pressure: 75kPa – 106kPa
mains voltage within ±10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2,000 meters
●
●
●
●
●
●