LIMITED WARRANTY AND LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and 

service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services 

are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke 

authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, 

has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation 

or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications 

for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software 

will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but 

have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if 

product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. 

Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased 

in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or 

replacement of a defective product which is returned to a Fluke authorized service center within the warranty 

period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization 

information, then send the product to that service center, with a description of the difficulty, postage and insurance 

prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will 

be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by 

neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including 

overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical 

components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. 

Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the 

repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER 

WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF 

MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY 

SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF 

DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation 

of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every 

buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of 

competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

11/99

Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.

Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands

Содержание 9102S

Страница 1: ...Well Calibrator User s Guide 2005 Rev 1 4 20 2005 2020 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective c...

Страница 2: ...nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB D...

Страница 3: ...on 5 DC Power Operation 5 Setup 5 Parts and Controls 6 Rear Panel 6 Front Panel 8 Accessories 9 General Operation 10 Set the Temperature 10 Change the Display Units 10 Operation 10 Well Temperature 12...

Страница 4: ...e 20 RS 232 Connection 21 Interface Commands 22 Test Probe Calibration 23 Calibrate a Single Probe 24 Dry well Characteristics 24 Stabilization and Accuracy 24 Calibration and Adjustment Procedure 25...

Страница 5: ...control Eight set point memory Adjustable readout in C or F The temperature is accurately controlled by Fluke Calibration s digital controller The controller uses a precision platinum RTD as a sensor...

Страница 6: ...uct visit http flukecal com register product Safety Information A Warning identifies conditions and procedures that are dangerous to the user A Caution identifies conditions and procedures that can ca...

Страница 7: ...11 Group 1 Class A Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequency energy that is necessary for the internal function of the equipment itself Class A Equ...

Страница 8: ...only For Maintenance and cleaning recommendations see Maintenance Quick Start Make sure to read the dry out period text in the Safety Information that comes with the Product Unpacking Unpack the Produ...

Страница 9: ...olarity as shown in Figure 1 The outside conductor is positive and the inside is negative The ac power switch on the rear panel does not switch the dc power Figure 1 12 V dc Power Source Polarity Setu...

Страница 10: ...Product The switch is either on or off The on position is for normal operation The off position disconnects power to the entire unit Fan The Product has a variable speed fan Under certain circumstanc...

Страница 11: ...Dry well Calibrator Parts and Controls 7 Figure 2 Back Panel...

Страница 12: ...the wells when tipped forward Inspect the strap periodically for wear Well Block Located on the middle of the front panel are the well openings where probes may be inserted into the well The block acc...

Страница 13: ...to store parameters to the displayed value d down arrow Used to decrement the displayed value of parameters u up arrow Used to increment the displayed value EXIT Used to exit a function and to skip to...

Страница 14: ...duct can display temperature in Celsius or Fahrenheit The temperature units are shipped from the factory set to Celsius There are two ways to change to Fahrenheit or back to Celsius 1 Press SET and u...

Страница 15: ...IT SET SET Select Setpoint Adjust Setpoint Units C F Scan On Off Scan Rate Display Temperature Configuration Menu Secondary Functions x 5 HL Adj HL SET EXIT Units C F Set Proportuinal Band Select R0 A...

Страница 16: ...y and adjusting the set point value Programmable Set points The controller stores 8 set point temperatures in memory The set points can be quickly recalled to conveniently set the Product to a previou...

Страница 17: ...t are ignored S Accept new set point value To change temperature units see Change the Display Units Scan The scan rate can be set and enabled so that when the set point is changed the dry well heats o...

Страница 18: ...t is recalculated when the set point temperature is changed The set point resistance is used to perform a calibration adjustment using the Callendar Van Dusen R versus T curve fit see Calibration and...

Страница 19: ...h as this the heater output power is proportional to the well temperature over a limited range of temperatures around the set point This range of temperature is called the proportional band At the bot...

Страница 20: ...he proportional band press u or d 10 0 New proportional band setting To accept the new setting press SET Press EXIT to continue without storing the new value S Accept the new proportional band setting...

Страница 21: ...RS 232 interface parameters menu is indicated by SErL Serial RS 232 interface parameters menu Press SET to enter the menu The serial interface parameters menu contains parameters which determine the...

Страница 22: ...ll duplex any commands received by the Product through the serial interface are immediately echoed or transmitted back to the device of origin With half duplex the commands are executed but not echoed...

Страница 23: ...parameters Hard Cutout R0 ALPHA and DELTA which characterize the resistance temperature relationship of the platinum control sensor These parameters may be adjusted to improve the accuracy of the Pro...

Страница 24: ...of this parameter is set at the factory for best Product accuracy DeLta Flashes for one second and then the DELTA setting displays 0 0000 Current DELTA setting To change the DELTA setting press u or d...

Страница 25: ...l for serial communications is 8 data bits 1 stop bit and no parity Use no flow control Set the linefeed to ON all carriage returns are followed by a linefeed LF ASCII 10 and the duplex to HALF disabl...

Страница 26: ...terminating carriage return CR ASCII 13 is implied with all commands Table 3 Controller Communications Commands Command Description Command Format Command Example Returned Returned Example Acceptable...

Страница 27: ...ff lf eed of f lf of Calibration Menu Read R0 calibration parameter r 0 r0 999 999 r0 100 7 Set R0 calibration parameter to n r 0 n r 100 7 95 0 to 105 0 Read ALPHA calibration parameter al pha al al...

Страница 28: ...such a way as to compensate for nominal heat losses out of the top of the dry well However actual heat losses vary with temperature and the design of the thermometer probes inserted into the Product...

Страница 29: ...just within the most useful operating range of the Product The farther apart the calibration temperatures the greater the calibrated temperature range However the calibration error is also greater ove...

Страница 30: ...SET and d at the same time where T1 and R1 are the measured temperature and set point resistance at 2 0 C T2 and R2 are the measured temperature and set point resistance at 50 0 C T3 and R3 are the me...

Страница 31: ...epeatability Compare the error observed at each calibration point with the Product accuracy specification to determine accuracy performance If the Product is adjusted repeat the calibration procedure...

Страница 32: ...s are wiped clean of any debris loosened in the buffing process Periodic cleaning of the outside of the inserts ensures easy insertion and removal of the inserts from the well Handle the Product with...

Страница 33: ...alibration Parameters Allow the Product to stabilize and verify the accuracy of the temperature reading Controller locked up The controller may have locked up due to a power surge or other aberration...

Страница 34: ...he Product was turned off at high temperatures and reenergized too quickly Turn off the Product until the display is completely off and then reenergize Display flickers when the Product is turned off...