8845A/8846A
Users Manual
D-
2
Table D-2. Ohms Analog Filter Errors
Range
NPLC
Additional Analog Filter Error
10, 100
0.5 m
Ω
10
Ω
<10 1.9
m
Ω
10, 100
1.5 m
Ω
100
Ω
<10 9.0
m
Ω
10, 100
0.6
Ω
100 k
Ω
<10 2.5
Ω
Table D-3. DC Current Analog Filter Errors
Range
NPLC
Additional Analog Filter Error
100
μ
A, 10 mA, 1 A
100
0.005 % of range
10
0.015 % of range
1
0.027 % of range
0.2
0.09 % of range
0.02
0.27 % of range
1 mA, 100 mA, 10 A
[1]
10
0.001 % of range
1
0.0025 % of range
0.2
0.009 % of range
0.02
0.026 % of range
[1]
3 A range uses the 10 A range errors.