Making Measurements
Measuring Resistance
4
4-9
0.2173
8
4
F1
F2
F3
F4
F5
caw031.eps
3.
If not already highlighted as shown above, press the
4WIRE
soft key to switch to a
four-wire measurement.
2
INPUT
SENSE
HI
LO
FRONT
REAR
10A
1V
400
mA
4W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
caw023.eps
Figure 4-2. Input Connections for 4-wire Resistance Measurements
To make a four-wire resistance measurement using Fluke’s 2X4 test leads:
1.
Connect the test leads to the Meter’s input connectors as show in Figure 4-3.
2.
Press
N
.
3.
If not already highlighted, press the
2X4WIRE
soft key.