Pressure Calibrator
Performance Verification Tests
9
Table 3. Pressure Verification Tests (cont.)
5000 PSI
Input
Pressure
Lower Limit
Upper Limit
5000.0 4998.2 5001.8
4000.0 3998.2 4001.8
3000.0 2998.2 3001.8
2000.0 1998.2 2001.8
1000.0 998.2 1001.8
0.0 -1.8 1.8
10. Carefully vent all pressure and disconnect the second pressure port.
DC Voltage Measure Verification
To do the dc voltage measure verification test:
1. Push
repeatedly until VOLTS is shown on the third row of the Product
display.
2. Connect the Product red banana jack (V mA) to the Calibrator HI VOLTS
output.
3. Connect the Product black banana jack (COM) to the Calibrator LO VOLTS
output.
4. Set the Calibrator for the voltage setting in Table 4, and verify the display
reading on the Product.
Table 4. DC Voltage Measure Verification Tests
Applied Voltage from 7526A
Lower limit
Upper Limit
0.000 V
-0.002 V
0.002 V
15.000 V
14.996 V
15.004 V
30.000 V
29.994 V
30.007 V
5. Set the Calibrator to Standby.
Содержание 721-1601
Страница 4: ...721 Calibration Manual ii...
Страница 6: ...721 Calibration Manual iv...
Страница 7: ...v List of Figures Figure Title Page 1 Battery Replacement 17 2 User Replaceable Parts 20...
Страница 8: ...721 Calibration Manual vi...
Страница 25: ...Pressure Calibrator Maintenance 17 hmq061 eps Figure 1 Battery Replacement...
Страница 28: ...721 Calibration Manual 20 1 6 11 13 3 5 2 4 10 9 12 8 7 14 15 hmq101 eps Figure 2 User Replaceable Parts...