721
Calibration Manual
6
Required Equipment
The equipment listed in Table 2 is necessary to do the performance verification
tests and calibration adjustment.
Table 2. Required Equipment for Verification and Calibration
Equipment
Minimum Specifications
Recommended Model
Low-thermal EMF banana to
banana leads
-
Pomona 2948-36-2 (red)
and 2948-36-0 (black)
RTD adapter
-
Fluke 720URTDA
Calibrator/Resistance Standard
0 V to 30 V
Accuracy: 0.006 % rdg
±
0.5 mV
0 mA to 24 mA
Accuracy: 0.006 % rdg
±
0.25
μ
A
-40
°
C to 150
°
C RTD
0.033
°
C
Fluke Calibration 7526A
Pressure Controller/Calibrator
-14 PSI to 1000 PSI
0.005 %
Ruska 7250xi
Piston Gauge
3000 PSI to 5000 PSI
0.005 %
Fluke Calibration PG7302 or
PG7202
Metrology Well
-30 °C to 140 °C
0.06 °C
Fluke Calibration 9170
Performance Verification Tests
Warning
To prevent possible electrical shock, fire, or personal injury, do
not perform the performance test or verification test procedures
unless the Product is fully assembled.
The performance tests verify the complete operation of the Product and measure
the accuracy of each function against Product specifications. If the Product fails a
part of the test, calibration adjustment or repair is necessary. See “Calibration
Adjustment Procedure”.
The performance verification tests check the accuracy of each Product function
against its specifications. If the Product fails any of these tests, calibration
adjustment or repair is necessary.
Содержание 721-1601
Страница 4: ...721 Calibration Manual ii...
Страница 6: ...721 Calibration Manual iv...
Страница 7: ...v List of Figures Figure Title Page 1 Battery Replacement 17 2 User Replaceable Parts 20...
Страница 8: ...721 Calibration Manual vi...
Страница 25: ...Pressure Calibrator Maintenance 17 hmq061 eps Figure 1 Battery Replacement...
Страница 28: ...721 Calibration Manual 20 1 6 11 13 3 5 2 4 10 9 12 8 7 14 15 hmq101 eps Figure 2 User Replaceable Parts...