Calibration Bath
Maintenance
53
figure 10.eps
Figure 10. Calibration Example
Maintenance
The calibration instrument has been designed with the utmost care. Ease of
operation and simplicity of maintenance have been a central theme in the
product development. Therefore, with proper care the instrument should require
very little maintenance. Avoid operating the instrument in dirty or dusty
environments.
•
A battery is used to maintain operating parameters in the unit. All operating
parameters, including calibration parameters should be checked on a regular
basis to insure accuracy and proper operation of the instrument. See the
troubleshooting section for the procedure on checking the status of the
battery.
•
If the outside of the bath becomes soiled, it may be wiped clean with a damp
cloth and mild detergent. Do not use harsh chemicals on the surface which
may damage the paint.
•
Periodically check the fluid level in the bath to ensure that the level has not
dropped. A drop in the fluid level affects the stability of the bath. Changes in
fluid level are dependent upon several factors specific to the environment in
which the equipment is used. A schedule cannot be outlined to meet each
environmental setting. Therefore, the first year the bath should be checked
weekly with notes kept as to changes in bath fluid. After the first year, the