Calibration Bath
Controller Operation
41
Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting
determines the serial communications transmission rate.
The baud rate parameter is indicated by,
Serial baud rate parameter
Press “SET” to choose to set the baud rate. The current baud rate value will then
be displayed.
Current baud rate
The baud rate of the bath serial communications may be programmed to
300,600,1200, or 2400 baud. Use “UP” or “DOWN” to change the baud rate
value.
New baud rate
Press “SET” to set the baud rate to the new value or “EXIT” to abort the
operation and skip to the next parameter in the menu.
Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature
measurements transmitted from the serial interface. If the sample rate is set to 5,
the bath transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. The sample period is indicated by,
Serial sample period parameter
Press “SET” to choose to set the sample period. The current sample period value
will be displayed.
Current sample period (seconds)
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample rate
to the displayed value.
New sample period
BAUd
1200 b
2400 b
SAmPLE
SA= 1
SA= 60