Страница 7: ...ed For information on recalibrating your 5623B contact Fluke for assistance see How to Contact Fluke Depending on the user s Quality Assurance requirements the PRT drift should be checked periodically...

Страница 8: ...periods of time can cause the handle to become hot Follow all safety guidelines listed in the Users Guide Calibration Equipment should only be used by trained personnel Use the Product only as specif...

Страница 9: ...Sheath Dimensions Length x Diameter 152 mm x 6 35 mm 6 in x 0 25 in Sheath Material Inconel 600 Transition Junction Temperature Range 2 100 C to 156 C Minimum Immersion Length 3 5 mK error 114 mm 4 5...

Страница 10: ...Guidelines Keep the thermometer as clean as possible Immerse the thermometer in the appropriate liquid for the temperature range If a dry block is used the well diameter should allow the PRT to comfor...

Страница 11: ...th required can be determined by performing a gradient test taking measurements approximately every 1 27 cm 0 5 in until there is a significant difference in readings Allow the thermometer to stabiliz...

Страница 12: ...ood condition and the connections are good the PRT may be damaged Contact Fluke for assistance see How to Contact Fluke If the data is unstable at high temperatures it may be due to electrical noise i...