SC300 Option
Calibration and Verification of Square Wave Functions
6
6-83
low frequency band, then the high frequency band for the second amplitude, and so on,
until the flatness calibration is complete.
Press the OPTIONS and NEXT SECTION blue softkeys until the display reads “Set up
to measure leveled sine flatness”.
6-109. Low Frequency Calibration
Connect the Calibrator Mainframe SCOPE connector to the 5790A WIDEBAND input
as described under “Equipment Setup for Low Frequency Flatness”.
Follow these steps to calibrate low frequency Leveled Sine Wave flatness for the
amplitude being calibrated:
1.
Press the GO ON blue softkey.
2.
Establish the 50 kHz reference:
•
Allow the 5790A rms reading to stabilize.
•
Press the 5790A Set Ref blue softkey. (Clear any previous reference by pressing
the 5790A Clear Ref blue softkey prior to setting the new reference if required.)
3.
Press the GO ON blue softkey.
4.
Adjust the amplitude using the Calibrator Mainframe front panel knob until the
5790A reference deviation matches the 50 kHz reference within 1000 ppm.
5.
Repeat steps 1 to 4 until the Calibrator Mainframe display indicates that the
reference frequency is now 10 MHz. Continue with the high frequency calibration.
6-110. High Frequency Calibration
Connect the Calibrator Mainframe SCOPE connector to the power meter and power
sensor as described in, “Equipment Setup for High Frequency Flatness” later in this
section.
Follow these steps to calibrate high frequency Leveled Sine Wave flatness for the
amplitude being calibrated.
1.
Press the GO ON blue softkey.
2.
Establish the 10 MHz reference:
•
Press the power meter SHIFT key, then FREQ key and use the arrow keys to
enter the power sensor’s 10 MHz Cal Factor. Ensure that the factor is correct,
then press the power meter ENTER key.
•
Allow the power meter reading to stabilize.
•
Press the Power meter REL key.
3.
Press the GO ON blue softkey.
4.
Press the power meter SHIFT key, then FREQ key and use the arrow keys to enter
the power sensor’s Cal Factor for the frequency displayed on the Calibrator
Mainframe. Ensure that the factor is correct, then press the power meter ENTER
key.
5.
Adjust the amplitude using the Calibrator Mainframe front panel knob until the
power sensor reading matches the 10 MHz reference within 0.1%.
6.
Repeat steps 1 to 5 until the Calibrator Mainframe display indicates that either the
reference frequency is now 50 kHz or that the next steps calibrate pulse width.
Repeat the low frequency calibration procedure for the next amplitude unless the
Содержание 5500A
Страница 12: ...5500A Service Manual x...
Страница 16: ...5500A Service Manual 1 2...
Страница 46: ...5500A Service Manual 1 32...
Страница 48: ...5500A Service Manual 2 2...
Страница 98: ...5500A Service Manual 4 2...
Страница 101: ...Maintenance Access Procedures 4 4 5 om016f eps Figure 4 1 Exploded View of Rear Panel Assemblies...
Страница 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Страница 116: ...5500A Service Manual 5 2...
Страница 120: ...5500A Service Manual 5 6 MP3 MP14 4X MP26 MP2 H13 X16 5500A Final Assembly 5 of 6 om018f eps Figure 5 1 Chassis Assembly...
Страница 127: ...6 1 Chapter 6 Oscilloscope Calibration Options Option 5500A SC600 see page 6 3 Option 5500A SC300 see page 6 65...
Страница 128: ...5500A Service Manual 6 2...
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ......
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 268: ......
Страница 269: ......
Страница 270: ......
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 277: ......
Страница 278: ......
Страница 279: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ......
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ...DS2E SL2 DC5V RELAY...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......