SC600 Option
Verification
6
6-33
6-49. Verification at 50
Ω
For the 50
Ω
verification, connect the Calibrator Mainframe’s SCOPE connector to the
HP 3458A input, using the cable supplied with the Calibrator Mainframe, the external
50
Ω
termination, and the BNC(f) to Double Banana adapter. (The 50
Ω
termination is
closest to the HP 3458A input.) Connect the Calibrator Mainframe TRIG OUT
connector to the HP 3458A Ext Trig connector located on the rear of that instrument.
Make sure the Calibrator Mainframe impedance is set to 50
Ω
. (The blue softkey under
Output @ toggles the impedance between 50
Ω
and 1 M
Ω
). Proceed with the following
steps:
1.
Set the HP 3458A to DCV, NPLC = .01, TRIG EXT, and the DELAY to .0007 for
measuring the topline of the wave form, and the DELAY to .0012 for measuring the
baseline of the wave form. Manually lock the HP 3458A to the range that gives the
most resolution for the topline measurements. Use this same range for the
corresponding baseline measurements at each step. See Table 6-22.
2.
Enable the Calibrator Mainframe external trigger by toggling the blue softkey under
TRIG to /1.
3.
Measure the topline first, as indicated in Table 6-22. For each measurement, take
samples for at least two seconds, using the HP 3458A MATH functions to determine
the average or mean value. See “Setup for SC600 Voltage Square Wave
Measurements” for more details.
4.
Measure the baseline of each output after the corresponding topline measurement, as
indicated in Table 6-22. The peak-to-peak value is the difference between the topline
and baseline measurements. Compare the result to the tolerance column.
Table 6-22. AC Voltage Verification at 50
Ω
Calibrator
Mainframe
Output
(1 kHz)
HP 3458A
Range
Topline
Reading
Baseline
Reading
Peak-to-Peak
Peak-to-Peak x
Correction
Tolerance
(
±
V)
1 mV
100 mV dc
0.000043
-1 mV
100 mV dc
0.000043
10 mV
100 mV dc
0.000065
-10 mV
100 mV dc
0.000065
25 mV
100 mV dc
0.000103
-25 mV
100 mV dc
0.000103
110 mV
100 mV dc
0.000315
-110 mV
100 mV dc
0.000315
500 mV
1 V dc
0.00129
-500 mV
1 V dc
0.00129
2.2 V
10 V dc
0.00554
-2.2 V
10 V dc
0.00554
6.6 V
10 V dc
0.01654
-6.6 V
10 V dc
0.01654
Содержание 5500A
Страница 12: ...5500A Service Manual x...
Страница 16: ...5500A Service Manual 1 2...
Страница 46: ...5500A Service Manual 1 32...
Страница 48: ...5500A Service Manual 2 2...
Страница 98: ...5500A Service Manual 4 2...
Страница 101: ...Maintenance Access Procedures 4 4 5 om016f eps Figure 4 1 Exploded View of Rear Panel Assemblies...
Страница 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Страница 116: ...5500A Service Manual 5 2...
Страница 120: ...5500A Service Manual 5 6 MP3 MP14 4X MP26 MP2 H13 X16 5500A Final Assembly 5 of 6 om018f eps Figure 5 1 Chassis Assembly...
Страница 127: ...6 1 Chapter 6 Oscilloscope Calibration Options Option 5500A SC600 see page 6 3 Option 5500A SC300 see page 6 65...
Страница 128: ...5500A Service Manual 6 2...
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ......
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 268: ......
Страница 269: ......
Страница 270: ......
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 277: ......
Страница 278: ......
Страница 279: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ......
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ...DS2E SL2 DC5V RELAY...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......