High Voltage Load Adapter

 

 

Performance and Verification Test

   

  

Note 

A Fluke 5520A or 5500A Calibrator is 

recommended as a source for the voltage 

method due to its output voltage/current 

capabilities. A Fluke 8845A Digital 

Multimeter (or equivalent) is 

recommended for measurements with 

either test method. 

For the voltage test method, test each resistor as 

follows: 
1.  Apply the recommended voltage at power 

mains frequency (either 50 or 60 Hz) 

between the terminal for the resistor being 

measured and the COM terminal. 

2.  Measure the current flowing through the load. 
3.  Calculate the resistance by dividing the 

sourced voltage by the measured current 

(R

L

=V

s

/I

m

). 

The calculated resistance should be within 10 % 

of the nominal resistance value found in Table 1. 
Repeat steps 1 through 3 for each Load resistor 

terminal, adjusting the applied voltage per 

Table 1. 
For the alternative multimeter test method, 

measure each resistor as follows: 
1.  Place the probes of a multimeter between the 

selected resistor input terminal and the COM 

terminal. 

2.  Read the measured resistor using the 

resistance mode of the multimeter, and note 

the value. 

3.  Insure the measured value is within 10 % of 

the nominal resistance. 

Repeat steps 1 through 3 for each Load resistor 

terminal per Table 1. 
 

POWER

I

O

0

1

2

3

4

5

6

7

8

9

ENTER

M

k

m

V

Hz

FIELD

EDIT

/

+

F

OPR

EARTH

EXGRD

SCOPE

MENU

PREV

SHIFT

RESET

CE

SETUP

REF

NEW

TC

MEAS

¡F

µ

n

p

W

dBm

sec

¡C

A

MULT

x

DIV

÷

MODES

MORE

STBY

HI

LO

TRIG

GUARD

TC

20A

NORMAL

AUX

5520A

 CALIBRATOR

SCOPE

OUT

V,   ,   ,RTD

A,    -SENSE, AUX V

20V PK MAX

20V PK MAX

CALIBRATION

ENABLE

SERIAL 1
FROM HOST

IEEE-488

IN                    OUT

 /

63 Hz

47 Hz
600VA MAX

FLUKE CORPORATION
MADE IN USA

PATENTS PENDING

NO INTERNAL USER SERVICEABLE
PARTS. REFER SERVICE TO 
QUALIFIED SERVICE PERSONNEL

NORMAL

WARNING:

TO AVOID ELECTRIC SHOCK GROUNDING 

CONNECTOR IN POWER CORD MUST BE CONNECTED

FUSE

CAUTION

FOR FIRE PROTECTION REPLACE ONLY

WITH A 250V FUSE OF INDICATED RATING

MAINS SUPPLY

100V

/

120V

220V

/

240V

T5.0A 250V (SB)
T2.5A 250V (SB)

CHASSIS
GROUND

SERIAL 2
TO UUT

10 MHz

MAX

5V P - P

WARNING:

TO AVOID PHYSICAL INJURY, INSURE THAT THE FILTER 

IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRUMENT

TO CLEAN THE FILTER:

-UNPLUG INSTRUMENT
-REMOVE FILTER
-FLUSH WITH SOAPY WATER
-DRY BEFORE REINSTALLATION 

INSTALLED
OPTIONS

        - SC300

        - SC600

5320A-LOAD

 

1200 V MAX
140 W MAX

2000 V MAX
110 W MAX

2000 V MAX

80 W  MAX

5500 V MAX
300 W MAX

5500 V MAX
120 W MAX

5500 V MAX

60 W MAX

5500 V MAX

30 W MAX

5500 V MAX

6 W  MAX

COM

10 k

35 k

500 k

1M

50 k

100 k

250 k

5 M

100 mA

40 mA

10 mA

4 mA

1mA

POWER

I

O

20 V MAX

PROTECTIVE

EARTH (GROUND)

POWER INPUT
12 V

10 VA MAX.

NO USER SERVICEABLE PARTS CONTAINED WITHIN.
HAZARDOUS VOLTAGES PRESENT. DO NOT REMOVE COVERS.

SAFETY WARNING

DISCONNECT POWER AND SIGNAL LEADS
BEFORE SERVICING.

READ MANUAL BEFORE USE.

REFER SERVICE TO QUALIFIED PERSONNEL.

FLUKE CORPORATION

www.fluke.com

DURING USE,

TERMINAL MUST BE 
CONNECTED TO THE 
PROTECTIVE EARTH 

OF THE  5320A

8846A

 

6-1/2 DIGIT  PRECISION MULTIMETER

0.20100

mA~

DC V

 ACV

DC I

AC I

TRIG

FREQ

PERIOD

TEMP

ZERO

INSTR

SETUP

MEAS

SETUP

RANGE

F1

F2

F3

F4

F5

BACK

MEMORY

ANALYZE

INPUT

SENSE

HI

LO

FRONT

 REAR

10A

1V

400

mA

4W

HI

LO

2W/4W

V

300V

FUSED

1000 V CAT I

   600V CAT II

Front

Rear

 

ehq104.eps 

Figure 4. Test Equipment Connections for Alternate Test Method 