background image

Instructions for use

MULTIflex LED coupling 460 LED - 1.007.3201

Содержание MULTIflex LED coupling 460 LED

Страница 1: ...Instructions for use MULTIflex LED coupling 460 LED 1 007 3201...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Страница 3: ...struments 14 5 4 Pulling the handpiece off 15 6 Checking for malfunctions and troubleshooting 16 6 1 Checking for malfunctions before initial start up 16 6 2 Troubleshooting 16 6 2 1 Replacing the LED...

Страница 4: ...ervice If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use are intended for...

Страница 5: ...erature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Do not dispose of with household wast...

Страница 6: ...use MULTIflex LED coupling 460 LED 1 007 3201 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to...

Страница 7: ...duct and accessories appropriately before disposal 2 2 Electrical shock Connecting a non KaVo system to the medical device can lead to electrical shock and injury to patient user and third parties 4 D...

Страница 8: ...vicing and safety checks may only be performed by trained service personnel The following persons are authorised to do this Service technicians of KaVo branches after the appropriate product training...

Страница 9: ...2 Safety 2 8 Disposal 9 28 2 8 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Страница 10: ...roduct must not be modified or used for any other purpose since this may be haz ardous The medical device is designed for coupling of the standard supply hose DIN EN ISO 9168 to MULTIflex compatible d...

Страница 11: ...the unit set the cold light intensity to the lowest intensity level and increase the brightness level maximally to intensity level 4 4 1065 1060 If the KaVo MULTIflex coupling is used on these units...

Страница 12: ...ge conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm up to a temperature of 20 C to 25 C 68 F to 77 F Temperat...

Страница 13: ...s the product and ac cessories WARNING Dispose of the product in appropriate manner Infection hazard 4 Reprocess the product and accessories before disposal Also refer to 7 Reprocessing steps in accor...

Страница 14: ...nut screw of the hose 4 Re tighten with the enclosed wrench 4 Apply a small amount of KaVo Spray to the O rings on the supply hose The coupling remains screwed onto the hose 5 2 Removing the coupling...

Страница 15: ...ED 1 007 3201 5 Operation 5 4 Pulling the handpiece off 15 28 4 Check the secure fit of the LUX instrument on the coupling by pulling on it 5 4 Pulling the handpiece off 4 Grasp the coupling and pull...

Страница 16: ...LED lamp by 180 and insert it Also refer to 6 2 1 Replacing the LED lamp Page 16 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and...

Страница 17: ...k properly Case 1 KaVo MULTI LED lamp is on Case 2 KaVo MULTI LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 KaVo MULTI LED lamp is r...

Страница 18: ...g 460 LED 1 007 3201 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting 18 28 4 Push the O ring to the front and remove it 4 Spray a lint free cloth with KaVo Spray apply it to the ne...

Страница 19: ...ctor only KaVo recommends the following products based on the compatibility of the ma terials The microbiological efficacy must be ensured by the disinfectant manu facturer and proven by an expert opi...

Страница 20: ...rocess it in a washer disinfector only NOTE Adapters are needed for automated cleaning Order adapter separately Also refer to 8 Components and miscellaneous Page 23 7 3 1 pre cleaning CAUTION Never re...

Страница 21: ...asher disinfector and the VARIO TD programme and a mild alkaline cleaner made by Dr Weigert In addition KaVo recommends the use of a rinsing agent 4 For programme settings as well as cleansers and dis...

Страница 22: ...steam steriliser after the steril isation cycle The medical device has a max temperature resistance of up to 138 o C 280 4 o F Sterilisation parameters Select a suitable process from the following st...

Страница 23: ...cialised dental dealers Components Description Material number KaVo MULTI LED lamp 1 007 5372 Miscellaneous Description Material number Key 0 411 1563 Spare washer 0 553 5262 O ring 6 65x0 8 1 004 277...

Страница 24: ...not be liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or con...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ...1 007 5973 bd 20230627 07 en...

Отзывы: