Instructions for use MULTIflex LED coupling 460 LED - 1.007.3201

2 Safety | 2.5 Accessories and combination with other equipment

8 / 28

To ensure optimum function and to prevent property damage, please comply
with the following instructions:

4

Service the medical device regularly with care products and systems as de-
scribed in the instructions for use.

4

The device should be reprocessed and stored in a dry location, according to
instructions, if it is not to be used for an extended period of time.

2.5 Accessories and combination with other equipment

Use of non-authorised accessories or non-authorised modifications of the device
could lead to injury.

4

Only use accessories that have been approved for combination with the
product by the manufacturer.

4

Only use accessories that are equipped with standardised interfaces.

4

Do not make any modifications to the product.

2.6 Qualification of personnel

Application of the product by users without the appropriate medical training
could injure patients, users or third parties.

4

Make sure that the user has read and fully comprehends the instructions for
use.

4

Make sure that the user has read and comprehends the national and re-
gional regulations.

4

The device may be used only if the user has completed the appropriate
medical training.

2.7 Service and repair

Repairs, servicing and safety checks may only be performed by trained service
personnel. The following persons are authorised to do this:

▪ Service technicians of KaVo branches after the appropriate product training

▪ Service technicians of KaVo authorised dealers after the appropriate product

training

Observe all the following items during servicing work:

4

Have the service and testing tasks carried out according to the Medical
Device Operator Ordinance.

4

KaVo recommends specifying in-house service intervals where the medical
device is brought to a professional shop for cleaning, servicing and a func-
tion check. Define the service interval depending on the frequency of use.

As a result of the use of NON-KaVo original spare parts during the repair, parts
such as covers may become undone and injure the patient, user or other
people. This may result in aspiration, swallowing of parts and possibly even a
risk of suffocation.

4

Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.

NOTE

If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.

The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by Medical Device Law §4, Abs.1 No. 1 and therefore requires its own
conformity check.

Содержание MULTIflex LED coupling 460 LED

Страница 1: ...Instructions for use MULTIflex LED coupling 460 LED 1 007 3201...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Страница 3: ...struments 14 5 4 Pulling the handpiece off 15 6 Checking for malfunctions and troubleshooting 16 6 1 Checking for malfunctions before initial start up 16 6 2 Troubleshooting 16 6 2 1 Replacing the LED...

Страница 4: ...ervice If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use are intended for...

Страница 5: ...erature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Do not dispose of with household wast...

Страница 6: ...use MULTIflex LED coupling 460 LED 1 007 3201 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to...

Страница 7: ...duct and accessories appropriately before disposal 2 2 Electrical shock Connecting a non KaVo system to the medical device can lead to electrical shock and injury to patient user and third parties 4 D...

Страница 8: ...vicing and safety checks may only be performed by trained service personnel The following persons are authorised to do this Service technicians of KaVo branches after the appropriate product training...

Страница 9: ...2 Safety 2 8 Disposal 9 28 2 8 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Страница 10: ...roduct must not be modified or used for any other purpose since this may be haz ardous The medical device is designed for coupling of the standard supply hose DIN EN ISO 9168 to MULTIflex compatible d...

Страница 11: ...the unit set the cold light intensity to the lowest intensity level and increase the brightness level maximally to intensity level 4 4 1065 1060 If the KaVo MULTIflex coupling is used on these units...

Страница 12: ...ge conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm up to a temperature of 20 C to 25 C 68 F to 77 F Temperat...

Страница 13: ...s the product and ac cessories WARNING Dispose of the product in appropriate manner Infection hazard 4 Reprocess the product and accessories before disposal Also refer to 7 Reprocessing steps in accor...

Страница 14: ...nut screw of the hose 4 Re tighten with the enclosed wrench 4 Apply a small amount of KaVo Spray to the O rings on the supply hose The coupling remains screwed onto the hose 5 2 Removing the coupling...

Страница 15: ...ED 1 007 3201 5 Operation 5 4 Pulling the handpiece off 15 28 4 Check the secure fit of the LUX instrument on the coupling by pulling on it 5 4 Pulling the handpiece off 4 Grasp the coupling and pull...

Страница 16: ...LED lamp by 180 and insert it Also refer to 6 2 1 Replacing the LED lamp Page 16 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and...

Страница 17: ...k properly Case 1 KaVo MULTI LED lamp is on Case 2 KaVo MULTI LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 KaVo MULTI LED lamp is r...

Страница 18: ...g 460 LED 1 007 3201 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting 18 28 4 Push the O ring to the front and remove it 4 Spray a lint free cloth with KaVo Spray apply it to the ne...

Страница 19: ...ctor only KaVo recommends the following products based on the compatibility of the ma terials The microbiological efficacy must be ensured by the disinfectant manu facturer and proven by an expert opi...

Страница 20: ...rocess it in a washer disinfector only NOTE Adapters are needed for automated cleaning Order adapter separately Also refer to 8 Components and miscellaneous Page 23 7 3 1 pre cleaning CAUTION Never re...

Страница 21: ...asher disinfector and the VARIO TD programme and a mild alkaline cleaner made by Dr Weigert In addition KaVo recommends the use of a rinsing agent 4 For programme settings as well as cleansers and dis...

Страница 22: ...steam steriliser after the steril isation cycle The medical device has a max temperature resistance of up to 138 o C 280 4 o F Sterilisation parameters Select a suitable process from the following st...

Страница 23: ...cialised dental dealers Components Description Material number KaVo MULTI LED lamp 1 007 5372 Miscellaneous Description Material number Key 0 411 1563 Spare washer 0 553 5262 O ring 6 65x0 8 1 004 277...

Страница 24: ...not be liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or con...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ...1 007 5973 bd 20230627 07 en...

Отзывы: