Maintenance and Troubleshooting
Downloading Recorded Data
8
8-5
SENSOR <n>: CALIBRATION INVALID
Cycle the power. Try a different sensor. Contact an
Authorized Service Center.
SENSOR <n>: CALIBRATION EXPIRED
Contact an Authorized Service Center to have the
sensor recalibrated. Check the time/date to make
sure it is correct.
POWER DOWN: CALCULATION ERROR
Contact an Authorized Service Center.
POWER DOWN: SOFTWARE RESET
Contact an Authorized Service Center.
POWER DOWN: HARDWARE RESET
Contact an Authorized Service Center.
BUTTONS: <button name>...
Check to make sure the buttons are not being
pressed. Cycle the power. Contact an Authorized
Service Center.
Start-up Error Messages
Error messages may appear after the self-test. If this happens, contact a Fluke Authorized
Service Center for assistance.
Downloading Recorded Data
The thermo-hygrometer stores up to 400,000 readings in its memory. These readings can
be downloaded to a computer using the RS-232 serial port.
The following steps should be followed to download logged data using the serial port:
1.
Connect the thermo-hygrometer to a COM port on the computer using the serial
cable.
2.
Run a terminal program such as Windows(r) Terminal or HyperTerminal(r).
a.
Set up the terminal program to use the correct COM port and baud rate.
Configure the communication settings to use 8 data bits, 1 stop bit, no parity
and no flow control.
b.
If you want to save the downloaded data to a file, configure your terminal
program to capture the incoming text and save it to a text file. Refer to your
terminal program’s documentation for details on how to do this.
3.
Begin downloading the data.
a.
Use the menu function DATA RECORD - DATA PRINT to begin sending
recorded data uning text format. See Data Print.
b.
The thermo-hygrometer should immediately begin sending the data. If no
readings have been stored, the thermo-hygrometer will not respond to these
commands.
4.
If the terminal program was set to capture the incoming data to a file, wait until
no more data is being received, then stop capturing.
The recorded data can also be transferred using the IR port with LogWare III software, or
written to a file with a data card.
Placement of IR Dongle for Transfer of Data
The thermo-hygrometer can be used with the LogWare III software to collect data via the
IR interface. Place your device near the IR port of the thermo-hygrometer between 3 and
6 inches (8 and 16 cm) and pointed directly at and perpendicular to the IR window.
Содержание 5020A
Страница 6: ...5020A Users Manual iv...
Страница 8: ...5020A Users Manual vi...
Страница 12: ...5020A Users Manual 1 2...
Страница 20: ...5020A Users Manual 2 2...
Страница 24: ...5020A Users Manual 3 2...
Страница 30: ...5020A Users Manual 3 8...
Страница 32: ...5020A Users Manual 4 2...
Страница 38: ...5020A Users Manual 5 2 Comm Setting 5 30 Serial 5 31 IR 5 32 Password 5 32 System Info 5 34...
Страница 72: ...5020A Users Manual 6 2...
Страница 124: ...5020A Users Manual 6 54...
Страница 126: ...5020A Users Manual 7 2...
Страница 130: ...5020A Users Manual 7 6...
Страница 132: ...5020A Users Manual 8 2...
Страница 138: ...5020A Users Manual 8 8...
Страница 141: ...Index continued 3 W warnings 1 4 Z zone types 5 9...
Страница 142: ...5020A Users Manual 4...