5020A
Users Manual
6-34
Response: 2003,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The response is returned in
the format, <year>,<month>,<day>.
CALibrate<chn>:DATE:CALibrate (<year>,<month>,<day>)
Sets the sensor calibration date for the specified channel.
Example: CAL1:DATE:CAL 2003,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <year>, <month>, and
<day> parameters must all be specified. The <year> parameter is a four-digit number.
The <month> parameter is a one or two-digit number (1 to 12). The <day> parameter is a
one or two-digit number (1 to 31). This command is password protected (see Password
Commands).
CALibrate<chn>:DATE:DUE?
Returns the sensor calibration due date for the specified channel.
Example: CAL1:DATE:DUE?
Response: 2004,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The response is returned in
the format, <year>,<month>,<day>.
CALibrate<chn>:DATE:DUE (<year>,<month>,<day>)
Sets the sensor calibration due date for the specified channel.
Example: CAL1:DATE:DUE 2004,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <year>, <month>, and
<day> parameters must all be specified. The <year> parameter is a four-digit number.
The <month> parameter is a one or two-digit number (1 to 12). The <day> parameter is a
one or two-digit number (1 to 31). This command is password protected (see Password
Commands).
CALibrate<chn>:EXPired?
Returns the sensor calibration expiration state for the specified channel.
Example: CAL1:EXP?
Response: 0
The CALibrate suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if
the sensor calibration has expired. A value of 0 is returned if the sensor calibration has
not expired.
CALibrate<chn>:PARameter:OFFSet<num>?
Returns the sensor calibration offset parameter for the specified channel.
Example: CAL1:PAR:OFFS2?
Response: -0.210
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The OFFSet suffix, <num>,
specifies either temperature (1) or humidity (2).
Содержание 5020A
Страница 6: ...5020A Users Manual iv...
Страница 8: ...5020A Users Manual vi...
Страница 12: ...5020A Users Manual 1 2...
Страница 20: ...5020A Users Manual 2 2...
Страница 24: ...5020A Users Manual 3 2...
Страница 30: ...5020A Users Manual 3 8...
Страница 32: ...5020A Users Manual 4 2...
Страница 38: ...5020A Users Manual 5 2 Comm Setting 5 30 Serial 5 31 IR 5 32 Password 5 32 System Info 5 34...
Страница 72: ...5020A Users Manual 6 2...
Страница 124: ...5020A Users Manual 6 54...
Страница 126: ...5020A Users Manual 7 2...
Страница 130: ...5020A Users Manual 7 6...
Страница 132: ...5020A Users Manual 8 2...
Страница 138: ...5020A Users Manual 8 8...
Страница 141: ...Index continued 3 W warnings 1 4 Z zone types 5 9...
Страница 142: ...5020A Users Manual 4...