Motor Analyzer

Motor Setup

5

Default values are available for the nameplate settings and ranges. Use 

 (UNIT SETUP) to open the 

setup screen to configure these values. The selection depends on the default motor frequency.
Softkeys:

Table 5. Motor Design Type and Characteristics Summary

Motor Design

NEMA-A

NEMA-B

NEMA-C

NEMA-D

NEMA-E

IEC-H

IEC-N

Starting Current

High

Medium

Medium

Medium

Medium

Medium

Medium

Starting Torque

Medium

Medium

High

Very High

Medium

High

Medium

Breakdown Torque

High

Medium

High

Very High

Medium

High

Medium

Table 6. Nameplate Settings and Ranges

Required from motor nameplate

Range

Resolution

Rated Power

746 W to 746 kW

1 W to 1 kW

1 hp to 1000 hp

1 hp

Rated Speed

 

Direct on Line

 

VSD

 

300 rpm to 3600 rpm

 

800 rpm to 3600 rpm

1 rpm

Rated Voltage

 

50 Hz motors

 

60 Hz motors

 

170 V to 1000 V

 

208 V to 1000 V

1 V

Rated Current

1 A to 1500 A

0.1 A

Rated Frequency

50 Hz or 60 Hz

- - -

Rated Cos Q (DPF)

0.01 to 1.00

0.01

Rated Service Factor

1.00 to 1.50

0.01

Motor Design Type

NEMA-A, NEMA-B, NEMA-C, NEMA-D, NEMA-E, IEC-H, 
IEC-N, Other

 - - -

Select UNIT SETUP screen

Opens the Motor Analyzer Function Preference screen. 
Select metric or imperial units for the mechanical motor 
power, torque, and the default motor frequency.

Select TREND SETUP screen

Adjusts the duration of the automatic trend recording. The 
default duration is 7 days.

Select DEFAULTS for the 
MOTOR SETUP screen

Sets the default values defined in the Function Preferences 
screen (see UNIT SETUP). The defaults depend on the 
default motor frequency. 

START measurements 

Starts the Motor Analyzer function. See 

Motor Analyzer 

Parameters

 on page 8.

Содержание 438-II

Страница 1: ...March 2016 Rev 1 10 17 2016 2017 Fluke Corporation All rights reserved All product names are trademarks of their respective companies 438 II Motor Analyzer Users Manual...

Страница 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Страница 3: ...Motor Measurements 3 Motor Setup 4 Unit Setup 6 Set Analyzer Limits 7 Trend Setup 7 Motor Analyzer Parameters 8 Mechanical Parameters 8 Electrical Parameters 9 View Analyzer Limit 9 Meter Screen 10 Tr...

Страница 4: ...ii 438 II Users Manual...

Страница 5: ...eed to use additional mechanical sensors or shut down the process The Motor Analyzer is for three wire three phase DELTA measurement configurations If any other configuration is active the feature pro...

Страница 6: ...oduct if it is damaged Disable the Product if it is damaged Do not work alone See the 434 II 435 II 437 II Users Manual for a complete list of Safety Warnings and explanation of Symbols used on the Pr...

Страница 7: ...to 100 0 1 3 1 NA Unbalance NEMA MG 1 0 to 100 0 1 0 15 5 Harmonics Voltage Factor NEMA MG 1 0 to 0 20 0 01 1 5 0 15 Unbalance Derating Factor 0 7 to 1 0 0 1 indicative NA Harmonics Derating Factor 0...

Страница 8: ...drives 5 Use to enter the motor nameplate parameters The Motor Analyzer supports motor designs according to NEMA and IEC design types See Table 5 and Table 6 If the design type is unknown select Othe...

Страница 9: ...ine VSD 300 rpm to 3600 rpm 800 rpm to 3600 rpm 1 rpm Rated Voltage 50 Hz motors 60 Hz motors 170 V to 1000 V 208 V to 1000 V 1 V Rated Current 1 A to 1500 A 0 1 A Rated Frequency 50 Hz or 60 Hz Rated...

Страница 10: ...7 Supported Variable Speed Drives Drive Characteristics Supported Range Drive Output Frequency 41 Hz to 69 Hz Drive Type Voltage Source Inverter only Drive Control Method V f Volt over Hertz only open...

Страница 11: ...Trend Setup screen is a simplified version of the normal TIMED START screen Motor Analyzer has an immediate start mode only and TIMED START is not available To open the screen go to Motor Setup screen...

Страница 12: ...wer supplied and mechanical power delivered in A colored bar graph indicates the performance of the motor compared to its rating The actual value for each parameter shows in the appropriate unit its r...

Страница 13: ...s View Analyzer Limit The View Analyzer Limit screen has the same information as the Set Analyzer Limits screen but is read only The only allowed action is OK or BACK to go to the Motor Analyzer scree...

Страница 14: ...ge 6 Speed rpm Efficiency Frequency Hz Harmonics HVF NEMA no unit Unbalance NEMA Harmonics Derating Factor HDF no unit Unbalance Derating Factor UDF no unit Total Derating Factor no unit The other par...

Страница 15: ...the 43X Series II that shows the recorded measurement data over time Note The Power Quality EVENT capture function is not available in Motor Analyzer to scroll through the METER screen Opens the ANAL...

Страница 16: ...alanced currents in the stator windings will result A small percentage of voltage unbalance will result in a much larger percentage of current unbalance Consequently the temperature rise of the motor...

Страница 17: ...ied When a voltage is applied to the motor that consists of components at frequencies other than the fundamental frequency the rated power of the motor should be derated according to the Derating Fact...

Страница 18: ...on the motor and entered in the Motor Setup screen The red area indicates the overload area for the motor A persistence function that plots the cross hair center in blue color is automatically enabled...

Отзывы: