Reference Pressure Gauge 

 

Serial Point Adjustment 

Calibration Steps 

To calibrate the Product, apply the values shown in Table 3 for the correct 

Product. If the Product fails to show the correct indication, adjustment is 

necessary. Calibration can be done in any pressure unit that can be shown on 

the Product display. Pressure units of psi and kPa are shown in Table 3 as 

examples. If calibrating in other units, use points that are as close as possible to 

those in Table 3. 

Table 3. Calibration Tests 

Model  

Calibration Points (psi) 

Calibration Points (kPa) 

2700G-BG100K 

0, -6, -12, -3, 0, 0, 3, 6, 9, 12, 15 

0, -40, -80, -20, 0, 0, 20, 40, 60, 80, 100 

2700G-BG200K 

0, -6, -12, -3, 0, 0, 6, 12, 18, 24, 30 

0, -40, -80, -20, 0, 0, 40, 80, 120, 160, 
200 

2700G-BG700K 

0, -6, -12, -3, 0, 0, 20, 40, 60, 80, 100 

0, -40, -80, -20, 0, 0, 140, 280, 420, 560, 
700 

2700G-BG2M 

0, -6, -12, -3, 0, 0, 60, 120, 180, 240, 300 

0, -40, -80, -20, 0, 0, 400, 800, 1200, 
1600, 2000 

2700G-BG3.5M 

0, -6, -12, -3, 0, 0, 100, 200, 300, 400, 500 

0, -40, -80, -20, 0, 0, 700, 1400, 2100, 
2800, 3500 

2700G-BG7M 

0, -6, -12, -3, 0, 0, 200, 400, 600, 800, 1000 

0, -40, -80, -20, 0, 0, 1400, 2800, 4200, 
5600, 7000 

2700G-G20M 

0, 600, 1200, 1800, 2400, 3000 

0, 4000, 8000, 12000, 16000, 20000 

2700G-G35M 

0, 1000, 2000, 3000, 4000, 5000 

0, 7000, 14000, 21000, 28000, 35000 

2700G-G70M 

0, 2000, 4000, 6000, 8000, 10000 

0, 14000, 28000, 42000, 56000, 70000 

Serial Point Adjustment  

Adjustment of the Product is done electronically with internal software. It is not 

necessary to open the Product case. All calibration commands and adjustments 

are done through the serial port connection. If the processes in the next sections 

do not bring the Product into specification, then it will be necessary to send your 

Product to Fluke Calibration for calibration adjustment. 

Set Zero 

After all connections are made, vent the Product to atmosphere and send this 

command: 
OFFSET_ADJ?  
Note the value that is shown. When the pressure is stable, send: 
OFFSET_ADJ N 
N is the pressure given from the previous OFFSET_ADJ? command. 

Set Span 

Send the GAIN_ADJ? command. 
Note the value shown. Use the correct pressure standard to input a value equal 

to or near the noted value. When pressure is stable, send: 
GAIN_ADJ N 

Содержание 2700G

Страница 1: ...pecifications are subject to change without notice All product names are trademarks of their respective companies 2700G Reference Pressure Gauge Calibration Manual Test Equipment Depot 800 517 8431 99...

Страница 2: ......

Страница 3: ...roduct 4 Change the Batteries 4 Specifications 6 Accuracy 6 Media Compatibility 6 Environmental Specifications 6 Mechanical Specifications 6 Available Pressure Ranges 7 Necessary Equipment 7 Calibrati...

Страница 4: ...2700G Calibration Manual ii...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Necessary Calibration Equipment 7 3 Calibration Tests 9 4 User Replaceable Parts 10...

Страница 6: ...2700G Calibration Manual iv...

Страница 7: ...v List of Figures Figure Title Page 1 Change the Batteries 5 2 Terminal Settings 8...

Страница 8: ...2700G Calibration Manual vi...

Страница 9: ...racy digital pressure test gauges Accurate to 0 02 FS the Product can be used as a calibration reference or in applications where high accuracy pressure measurement is necessary The Product features u...

Страница 10: ...safety procedures High pressure liquids and gases are hazardous and the energy from them can be released without warning To prevent possible electrical shock fire or personal injury Read all safety In...

Страница 11: ...information See manual Conforms to European Union directives Hazardous voltage Risk of electrical shock Conforms to relevant North American Safety Standards Conforms to relevant Australian standards T...

Страница 12: ...is not used for an extended period Be sure that the battery polarity is correct to prevent battery leakage Have an approved technician repair the Product Change the Batteries Warning To prevent possi...

Страница 13: ...Reference Pressure Gauge Maintenance 5 gsn002 eps Figure 1 Change the Batteries...

Страница 14: ...non toxic non explosive non oxidizing liquid or gas compatible with 316 stainless steel Environmental Specifications Operating Temperature 0 C to 50 C 32 F to 122 F Storage 20 C to 70 C 4 F to 158 F...

Страница 15: ...sary for calibration and adjustment To maintain the specified Product accuracy use a pressure standard with an uncertainty of 0 0075 FS of the Product or lower The necessary equipment for the calibrat...

Страница 16: ...l invalidate the compensation program in the Product For optimum results the calibration or adjustment should be done with ambient temperature as close as possible to 23 C 72 F Connections The Product...

Страница 17: ...00 3500 2700G BG7M 0 6 12 3 0 0 200 400 600 800 1000 0 40 80 20 0 0 1400 2800 4200 5600 7000 2700G G20M 0 600 1200 1800 2400 3000 0 4000 8000 12000 16000 20000 2700G G35M 0 1000 2000 3000 4000 5000 0...

Страница 18: ...e items and their prices contact a Fluke Calibration representative See the Contact Fluke Calibration section Table 4 User Replaceable Parts Description Fluke Part Number Gauge Cover 4201079 2700G 800...

Отзывы: