Calibration Manual 

Calibration Tests 

Calibration verifies the complete operation of the Product and measures the 

accuracy of each function against Product specifications. If the Product fails a 

part of the test, adjustment or repair is necessary. See “Serial Point Adjustment”. 
It is recommended that you apply full-scale pressure to the Product and then vent 

before calibration adjustment. Let the Product stabilize for 1 minute after venting 

before continued testing. 

Note 

Calibration and adjustment can be performed anywhere in the 
operating ambient temperature range of 18 

°

C to 28 

°

C (64 

°

F to 

82 

°

F). Calibration or adjustment outside this temperature range will 

invalidate the compensation program in the Product. For optimum 
results, the calibration or adjustment should be done with ambient 
temperature as close as possible to 23 

°

C (72 

°

F). 

Connections 

The Product uses a ¼ inch NPT male connection in the pressure input port. 

Adapters may be necessary to connect to the pressure standard. Make sure the 

hose, tubing, and fittings have a rated working pressure at or above the pressure 

of the unit. Make sure there are no leaks when you do the calibration process. 

Use Teflon tape where necessary.  

Initiate Communication 

Serial communications can be set up with terminal communication software on a 

PC, such as Windows HyperTerminal.  
1.  Connect the USB RS232 cable to the serial jack on the rear of the Product.  
2.  Connect the other end of the cable to the PC. Terminal settings are shown in 

Figure 2. 

 

hmo001.bmp 

Figure 2. Terminal Settings 

Содержание 2700G

Страница 1: ...pecifications are subject to change without notice All product names are trademarks of their respective companies 2700G Reference Pressure Gauge Calibration Manual Test Equipment Depot 800 517 8431 99...

Страница 2: ......

Страница 3: ...roduct 4 Change the Batteries 4 Specifications 6 Accuracy 6 Media Compatibility 6 Environmental Specifications 6 Mechanical Specifications 6 Available Pressure Ranges 7 Necessary Equipment 7 Calibrati...

Страница 4: ...2700G Calibration Manual ii...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Necessary Calibration Equipment 7 3 Calibration Tests 9 4 User Replaceable Parts 10...

Страница 6: ...2700G Calibration Manual iv...

Страница 7: ...v List of Figures Figure Title Page 1 Change the Batteries 5 2 Terminal Settings 8...

Страница 8: ...2700G Calibration Manual vi...

Страница 9: ...racy digital pressure test gauges Accurate to 0 02 FS the Product can be used as a calibration reference or in applications where high accuracy pressure measurement is necessary The Product features u...

Страница 10: ...safety procedures High pressure liquids and gases are hazardous and the energy from them can be released without warning To prevent possible electrical shock fire or personal injury Read all safety In...

Страница 11: ...information See manual Conforms to European Union directives Hazardous voltage Risk of electrical shock Conforms to relevant North American Safety Standards Conforms to relevant Australian standards T...

Страница 12: ...is not used for an extended period Be sure that the battery polarity is correct to prevent battery leakage Have an approved technician repair the Product Change the Batteries Warning To prevent possi...

Страница 13: ...Reference Pressure Gauge Maintenance 5 gsn002 eps Figure 1 Change the Batteries...

Страница 14: ...non toxic non explosive non oxidizing liquid or gas compatible with 316 stainless steel Environmental Specifications Operating Temperature 0 C to 50 C 32 F to 122 F Storage 20 C to 70 C 4 F to 158 F...

Страница 15: ...sary for calibration and adjustment To maintain the specified Product accuracy use a pressure standard with an uncertainty of 0 0075 FS of the Product or lower The necessary equipment for the calibrat...

Страница 16: ...l invalidate the compensation program in the Product For optimum results the calibration or adjustment should be done with ambient temperature as close as possible to 23 C 72 F Connections The Product...

Страница 17: ...00 3500 2700G BG7M 0 6 12 3 0 0 200 400 600 800 1000 0 40 80 20 0 0 1400 2800 4200 5600 7000 2700G G20M 0 600 1200 1800 2400 3000 0 4000 8000 12000 16000 20000 2700G G35M 0 1000 2000 3000 4000 5000 0...

Страница 18: ...e items and their prices contact a Fluke Calibration representative See the Contact Fluke Calibration section Table 4 User Replaceable Parts Description Fluke Part Number Gauge Cover 4201079 2700G 800...

Отзывы: