Calibration Adjustment
Save Calibration Data and Exit
5
5-43
Save Calibration Data and Exit
To save the calibration data and exit the Maintenance mode:
1. Remove all test leads from the Test Tool inputs.
2.
Press
F4
(
EXIT)
. The Test Tool shows on the display:
Calibration data valid.
Save data and exit maintenance mode?
Note
Calibration data valid indicates that the calibration adjustment
procedure is performed correctly. It does not necessarily mean that
the Test Tool meets the specifications listed in Chapter 2.
3. Press
F4
(
YES)
to save and exit.
Note
After saving the calibration data, the calibration number and date updates if
the calibration data changes and the data are valid.
The calibration number and date do not change if:
•
the calibration mode is entered and left without doing a calibration
adjustment.
•
only the contrast calibration adjustment and/or the probe calibration are
done.
If you press
F3
(
NO)
, the Test Tool returns to the calibration mode.
You can either calibrate the Test Tool again, or press
F4
(
EXIT)
,
F4
(
YES)
to save and exit.
Error messages:
WARNING: Calibration data not valid.
Save data and exit maintenance mode?
If you did the Warming-Up and Pre-Calibration successfully and you want to
store the Pre-Calibration data before continuing with the Final Calibration:
1. Press
F4
(YES).
When you turn off and turn on the Test Tool again, it will show the message:
The instrument needs calibration.
Please contact your service center.
The calibration date and number do not update. You must continue with the
Final Calibration.
To return to the Maintenance mode and repeat the complete calibration:
1. Press
F3
(NO).
2. Press
F1
until the display shows
WarmingUp (CL 0200):IDLE
, and
calibrate the Test Tool, starting at
Warming-Up and Pre-Calibration
.
If you want to exit and maintain the old calibration data:
1. Turn the Test Tool off.