Calibration Adjustment
Final Calibration (Firmware: V10.9 and Lower)
5
5-39
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points of Table 5-18.
8. Set the 5502A to Standby and continue at the
Input ABCD (AB) Zero
section.
Table 5-18. Input ABCD Gain Calibration Points
Cal step
UUT input value (5502A NORMAL)
CL 0799
5 mV
CL 0800
12.5 mV
CL 0801
25 mV
CL 0802
50 mV
CL 0803
125 mV
CL 0804
250 mV
CL 0805
500 mV
CL 0806
1.25 V
CL 0807
2.5 V
CL 0808
5 V
CL 0809
12.5 V
CL 0810
25 V
CL 0811
50 V (set 5502A to OPR!)
CL 0812
125 V
CL 0813
250 V
Input ABCD (AB) Zero
To do the Input ABCD (AB) Zero calibration:
1. Press
F2
to select calibration adjustment step CL 0852.
2. Short circuit Input A, B, C, D (A, B) with 50
Ω
feed through terminations.
3. Press
F3
to start the zero calibration.
4. Wait until the display shows the status
:READY
.
5. Remove the input terminations.
6. For Fluke 190-062, 190-102, and 190-202: continue at the
Multimeter (DMM)
Volt Gain
section. For Fluke 190-104 and 190-204: continue at the
Save
Calibration Data and Exit
section.