background image

 

Stik Thermometer 

 

Calibration 

Recommended Equipment 

The equipment listed in Table 1 will satisfy the calibration requirements: 

Table 1. Recommended Equipment 

Equipment 

Recommended Model 

Temperature Baths

 

Fluke 6331 40 

°

C to 300 

°

C Hot Bath 

Fluke 7381 -80 

°

C to 110 

°

C Cold Bath 

Fluke 7196 Equilibration Block 

Reference Thermometer

 

Fluke 5699-S Extended Range Metal-Sheath SPRT 

Reference Thermometer Readout 

1594A Super-Thermometer readout 

Computer 
 

Computer with an RS-232 com port 
RS-232 serial cable 
Terminal software to communicate with the 1551A 

Ex/1552A Ex Thermometer 

Initial Setup 

Check the batteries in the Product and replace them if necessary. The battery level 

indication is found in the Setup menu. 
Several settings in the Setup menu relate to measurement and should be set as follows:  

 

Sample Rate: 

 

1 per second 

 

Damping:   

 

OFF 

 

Display Resolution:  

0.001 

 

Auto Power Off: 

 

OFF 

The unit of measurement in the 1551A Ex/1552A Ex Thermometer may be set to °C or 

°F, but must match the unit of the reference thermometer. 
The baud rate in the computer software must match the baud rate of the 1551A Ex/1552A 

Ex Thermometer, which is usually 9600 but may be set to 2400. 

Содержание 1551A Ex

Страница 1: ...19 2010 2019 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 1551A Ex 1552A Ex Stik Thermomete...

Страница 2: ...Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destinati...

Страница 3: ...ial Setup 5 Record as found Calibration Parameters 6 Evaluate as found Measurement Errors 6 Adjustment at the Low Calibration Point 7 Adjustment at the Zero Calibration Point 8 Adjustment at the High...

Страница 4: ...1551A Ex 1552A Ex Calibration Manual ii...

Страница 5: ...tructions see the 1551A Ex 1552A Ex Users Manual How to Contact Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repai...

Страница 6: ...to Fluke Corporation Refer to How to Contact Fluke Substitution of components will impair suitability for hazardous locations Ex Safety Information This manual contains data and safety regulations th...

Страница 7: ...omagnetic Compatibility International IEC 61326 1 Portable Equipment IEC 61326 2 2 CISPR 11 Group 1 Class B Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequen...

Страница 8: ...a temperature bath having a known temperature and observing the difference between the temperature measured by the Product and the known temperature of the bath The temperature of the bath is determi...

Страница 9: ...RS 232 serial cable Terminal software to communicate with the 1551A Ex 1552A Ex Thermometer Initial Setup Check the batteries in the Product and replace them if necessary The battery level indication...

Страница 10: ...L USER TEMP1 ADJ1 CAL USER ADJ1 TEMP2 CAL USER TEMP2 ADJ2 CAL USER ADJ2 TEMP3 CAL USER TEMP3 ADJ3 CAL USER ADJ3 DATE CAL DEV DATE Evaluate as found Measurement Errors The Product should be tested to e...

Страница 11: ...stment Temperature Point 1551A Ex 1552A Ex Low 50 C 80 C Before adjustments to any calibration parameters can be made access to the protected calibration parameters must be enabled by sending the foll...

Страница 12: ...ameter to cause the Product to accurately measure the given temperature This adjustment will not affect the measurements at the other Adjustment temperatures Allow the readings of the Product to settl...

Страница 13: ...Protection will always be restored when the Product is switched on Evaluate as left Measurement Errors After all adjustments have been made the Product should be subjected to a final test to ensure t...

Страница 14: ...ent WWarning To prevent possible explosion fire or personal injury Change the batteries only in areas that are not Ex hazardous Replace the batteries when the low battery indicator shows to prevent in...

Страница 15: ...teries 3 Replace only with approved AAA batteries listed in this document See Table 10 Any substitution voids the Product safety rating 4 Make sure the battery polarity is correct 5 Replace the batter...

Страница 16: ...1551A Ex 1552A Ex Calibration Manual 12...

Отзывы: