background image

 

Stik Thermometer 

 

Calibration 

At each calibration point the measurement error may be calculated by subtracting the 

reference thermometer temperature from the 1551A Ex/1552A Ex temperature. The 

temperatures and measurement errors may be recorded in Table 4. 

Table 4. As-found Measurement Errors 

Point 

Reference (fill in) 

1551A Ex/1552A Ex (fill in) 

Error (fill in) 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
If the measurement errors are well within the specified accuracy, it may not be necessary 

to adjust the calibration parameters. If the error is larger than desired (say 50 % of the 

specified accuracy), then adjustments to the Product calibration parameters are necessary. 

Proceed to the next section. 

Adjustment at the Low Calibration Point 

When necessary, the first adjustment is made for the low end of the temperature range of 

the Product. Set the bath temperature to the Low Adjustment Temperature given in  

Table 5, depending on the model. Make sure the temperatures of the bath, reference 

thermometer, and Product are settled and stable before proceeding. 

Table 5. Low-Adjustment Temperature 

Point 

1551A Ex 

1552A Ex 

Low 

50 

°

80 

°

 
Before adjustments to any calibration parameters can be made, access to the protected 

calibration parameters must be enabled by sending the following command to the Product 

through the serial interface: 

SYST:PASS:CEN  <password>   

The parameter <password> is the current password, originally “1234”. 
To make an automatic adjustment at the low calibration point, issue the following 

command: 

CAL:USER:LOW  <temperature> 

The parameter <temperature> is the temperature of the bath as measured by the reference 

thermometer. Product accepts a temperature value with up to three decimal places of 

resolution. The temperature value must be in the same unit in which the 1551A 

Ex/1552A Ex is measuring. 
The command will set the value of the CAL:USER:TEMP1 parameter to the given 

temperature and automatically adjust the CAL:USER:ADJ1 parameter to cause the 

Product to measure the given temperature accurately. The adjustment made here will not 

affect the measurements at the other Adjustment temperatures. 
Allow the readings of the Product to settle and then verify that the readings match the 

temperature of the reference thermometer and are well within the specified accuracy of 

the 1551A Ex/1552A Ex. 

Содержание 1551A Ex

Страница 1: ...19 2010 2019 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 1551A Ex 1552A Ex Stik Thermomete...

Страница 2: ...Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destinati...

Страница 3: ...ial Setup 5 Record as found Calibration Parameters 6 Evaluate as found Measurement Errors 6 Adjustment at the Low Calibration Point 7 Adjustment at the Zero Calibration Point 8 Adjustment at the High...

Страница 4: ...1551A Ex 1552A Ex Calibration Manual ii...

Страница 5: ...tructions see the 1551A Ex 1552A Ex Users Manual How to Contact Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repai...

Страница 6: ...to Fluke Corporation Refer to How to Contact Fluke Substitution of components will impair suitability for hazardous locations Ex Safety Information This manual contains data and safety regulations th...

Страница 7: ...omagnetic Compatibility International IEC 61326 1 Portable Equipment IEC 61326 2 2 CISPR 11 Group 1 Class B Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequen...

Страница 8: ...a temperature bath having a known temperature and observing the difference between the temperature measured by the Product and the known temperature of the bath The temperature of the bath is determi...

Страница 9: ...RS 232 serial cable Terminal software to communicate with the 1551A Ex 1552A Ex Thermometer Initial Setup Check the batteries in the Product and replace them if necessary The battery level indication...

Страница 10: ...L USER TEMP1 ADJ1 CAL USER ADJ1 TEMP2 CAL USER TEMP2 ADJ2 CAL USER ADJ2 TEMP3 CAL USER TEMP3 ADJ3 CAL USER ADJ3 DATE CAL DEV DATE Evaluate as found Measurement Errors The Product should be tested to e...

Страница 11: ...stment Temperature Point 1551A Ex 1552A Ex Low 50 C 80 C Before adjustments to any calibration parameters can be made access to the protected calibration parameters must be enabled by sending the foll...

Страница 12: ...ameter to cause the Product to accurately measure the given temperature This adjustment will not affect the measurements at the other Adjustment temperatures Allow the readings of the Product to settl...

Страница 13: ...Protection will always be restored when the Product is switched on Evaluate as left Measurement Errors After all adjustments have been made the Product should be subjected to a final test to ensure t...

Страница 14: ...ent WWarning To prevent possible explosion fire or personal injury Change the batteries only in areas that are not Ex hazardous Replace the batteries when the low battery indicator shows to prevent in...

Страница 15: ...teries 3 Replace only with approved AAA batteries listed in this document See Table 10 Any substitution voids the Product safety rating 4 Make sure the battery polarity is correct 5 Replace the batter...

Страница 16: ...1551A Ex 1552A Ex Calibration Manual 12...

Отзывы: