Stik Thermometer 

 

Calibration 

The command will set the value of the CAL:USER:TEMP3 parameter to the given 

temperature and automatically adjust the CAL:USER:ADJ3 parameter to cause the 

Product to measure the given temperature accurately. The adjustment made here will not 

affect the measurements at the other Adjustment temperatures. 
Allow the readings of the Product to settle and then verify that the readings match the 

temperature of the reference thermometer well within the specified accuracy of the 

1551A Ex/1552A Ex. 

Set the Calibration Date 

The date of calibration should be stored in the Product using the following command: 

CAL:DEV:DATE  <year>,<month>,<day> 

Parameters <year>, <month>, and <day> are numbers representing the year, month, and 

day of the month respectively. 
After all the adjustments have been made and the calibration date has been set, protection 

of the calibration parameters should be restored using the following command: 

SYST:PASS:CDIS 

(Protection will always be restored when the Product is switched on.) 

Evaluate as-left Measurement Errors 

After all adjustments have been made, the Product should be subjected to a final test to 

ensure that it is measuring accurately over its entire range. The measurements are 

collected and errors are evaluated in the same way that was explained in “

Evaluating As-

found Measurement Errors

.” Table 8 may be used to record the as-left data. 

Table 8. As-left Measurement Errors 

Point 

Reference (fill in) 

1551A Ex/1552A Ex (fill in) 

Error (fill in) 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Record as-left Calibration Parameters 

As the last step in the calibration procedure, the final values of the Product calibration parameters 
should be recorded. The commands required to read the values of the calibration parameters are 
given in Table 9. Each command must be terminated with a carriage return character. The table 
may also be used to record the values of the calibration parameters. 

Table 9. As-left Calibration Parameters 

Parameter 

Command 

Value (fill in) 

TEMP1 

CAL:USER:TEMP1? 

 

ADJ1 

CAL:USER:ADJ1? 

 

TEMP2 

CAL:USER:TEMP2? 

 

ADJ2 

CAL:USER:ADJ2? 

 

TEMP3 

CAL:USER:TEMP3? 

 

ADJ3 

CAL:USER:ADJ3? 

 

DATE 

CAL:DEV:DATE? 

 

Содержание 1551A Ex

Страница 1: ...19 2010 2019 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 1551A Ex 1552A Ex Stik Thermomete...

Страница 2: ...Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destinati...

Страница 3: ...ial Setup 5 Record as found Calibration Parameters 6 Evaluate as found Measurement Errors 6 Adjustment at the Low Calibration Point 7 Adjustment at the Zero Calibration Point 8 Adjustment at the High...

Страница 4: ...1551A Ex 1552A Ex Calibration Manual ii...

Страница 5: ...tructions see the 1551A Ex 1552A Ex Users Manual How to Contact Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repai...

Страница 6: ...to Fluke Corporation Refer to How to Contact Fluke Substitution of components will impair suitability for hazardous locations Ex Safety Information This manual contains data and safety regulations th...

Страница 7: ...omagnetic Compatibility International IEC 61326 1 Portable Equipment IEC 61326 2 2 CISPR 11 Group 1 Class B Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequen...

Страница 8: ...a temperature bath having a known temperature and observing the difference between the temperature measured by the Product and the known temperature of the bath The temperature of the bath is determi...

Страница 9: ...RS 232 serial cable Terminal software to communicate with the 1551A Ex 1552A Ex Thermometer Initial Setup Check the batteries in the Product and replace them if necessary The battery level indication...

Страница 10: ...L USER TEMP1 ADJ1 CAL USER ADJ1 TEMP2 CAL USER TEMP2 ADJ2 CAL USER ADJ2 TEMP3 CAL USER TEMP3 ADJ3 CAL USER ADJ3 DATE CAL DEV DATE Evaluate as found Measurement Errors The Product should be tested to e...

Страница 11: ...stment Temperature Point 1551A Ex 1552A Ex Low 50 C 80 C Before adjustments to any calibration parameters can be made access to the protected calibration parameters must be enabled by sending the foll...

Страница 12: ...ameter to cause the Product to accurately measure the given temperature This adjustment will not affect the measurements at the other Adjustment temperatures Allow the readings of the Product to settl...

Страница 13: ...Protection will always be restored when the Product is switched on Evaluate as left Measurement Errors After all adjustments have been made the Product should be subjected to a final test to ensure t...

Страница 14: ...ent WWarning To prevent possible explosion fire or personal injury Change the batteries only in areas that are not Ex hazardous Replace the batteries when the low battery indicator shows to prevent in...

Страница 15: ...teries 3 Replace only with approved AAA batteries listed in this document See Table 10 Any substitution voids the Product safety rating 4 Make sure the battery polarity is correct 5 Replace the batter...

Страница 16: ...1551A Ex 1552A Ex Calibration Manual 12...

Отзывы: