Industrial ScopeMeter®
Test Tool Setup
11
Measurement Connections
The Test Tool has two 4 mm safety-shielded banana jack
inputs (red input A and blue input B) and a black safety
4 mm banana jack input (COM). See Figure 4.
hxv05.eps
Figure 4. Input Connections for Measurement
Input A
Use the input A (red) for all single input measurements.
Input B
For measurements on two different signals, use B input
(blue) together with input A (red).
COM
Use the COM input (black) as the single ground for low
frequency, continuity, Ohm (
Ω
), diode, and capacitance
measurements.
XW
Warning
To prevent electrical shock or fire, use only
one COM
˜
(common) connection, or ensure
that all connections to COM
˜
are at the
same potential.
Measurement Probe Setup
The Test Tool supports measurement probes, for
example:
•
10:1 voltage probe
•
1
mV/
°
C temperature probe
•
10 mV/A current clamp
To set up a specific probe type:
1. Push
to open the Scope and Meter menu.
2. Push
3
to open the
INPUT A
menu.
3. Use
Y
/
Z
to highlight
SELECT
.
4. Push
to open the
PROBE
select menu.
5. Use
Y
/
Z
to highlight the probe type.
6. Push
to accept the probe type and close the
menu.