Industrial ScopeMeter®
Safety Information
3
•
Do not use the Product if it operates
incorrectly.
•
Examine the case before you use the
Product. Look for cracks or missing
plastic. Carefully look at the insulation
around the terminals.
•
Do not use test leads if they are
damaged. Examine the test leads for
damaged insulation, exposed metal, or if
the wear indicator shows. Check test
lead continuity.
•
Use only cables with correct voltage
ratings.
•
Connect the common test lead before
the live test lead and remove the live test
lead before the common test lead.
•
Keep fingers behind the finger guards on
the probes.
•
Remove all probes, test leads, and
accessories before the battery door is
opened.
•
Remove all probes, test leads, and
accessories that are not necessary for
the measurement.
•
Do not exceed the Measurement
Category (CAT) rating of the lowest rated
individual component of a Product,
probe, or accessory.
•
Do not use a current measurement as an
indication that a circuit is safe to touch.
A voltage measurement is necessary to
know if a circuit is hazardous.
•
Disable the Product if it is damaged.
•
Do not use the Product if it is damaged.
•
Do not use the Product above its rated
frequency.
•
Do not use the Current Probe if it has
damaged insulation, exposed metal, or if
the wear indicator is visible.
•
Do not wear loose-fitting clothing or
jewelry and keep long hair tied back
when near rotating machinery. Use
approved eye protection and approved
personal-protective equipment where
necessary.