background image

 

PN 

4811274

 June 2016  Rev. 1, 11/18  

©2016-2018 Fluke Corporation. All rights reserved.

 

  1 

Y9600 

Rack-Mount Kit 

Instructions 

Introduction 

This sheet supplies instructions to mount the Product 
into a standard 19-inch equipment rack. The Rack-
Mount Slide Kit (the kit) contains all of the 
components necessary. 

W

Warning 

To prevent possible injury, do not restrict 
access to the Product power cord, which is 
the mains disconnecting device. If access 
to the mains power cord is inhibited by 
rack mounting, a properly-rated accessible 
proximate mains disconnecting switch 
must be provided as part of the 
installation. 

Note 

Review the mounting instructions provided by 

the manufacturer of the equipment rack before 

you proceed with the installation. 

Install Slides on the Product 

Install the slides on the Product before you install the 
Product into the rack. The numbers in Figure 1 show 
the listed process steps. 

Note 

Do not remove the top or bottom cover to rack-

mount the Product. 

 

To install the slides: 

1.  Push a flat-blade screwdriver between the 

chassis and the plastic strip to remove the plastic 
side-trim from the Product. Pry outward until the 
strip is free of the two grooves, and then pull the 
strip free from the Product.  

2.  Attach each rack-mount ear to the front of the 

Product: 

Note 

So that the bezel is not 

detached, complete one side 

before you start on the other 

side. 

a.  Remove the top and bottom hex screws from 

the side of each front handle. 

b.  Position each ear so that the holes in its tabs 

align with the screw holes on a front handle.  

c.  Use two M4 x 16mm or two 8-32 x ½ inch 

pan-head screws to attach each ear to the 
Product. See Figure 2 for screw identification.  

3.  Use the two 8-32 x ¼ inch truss-head screws to 

secure each of the two rack-mount filler bars to 
the Product. 

4.  Attach a slide bar to both sides of the Product. 

Secure each with three 8-32 x ¼ inch truss-head 
screws. 

5.  Remove the four plastic molded feet from the 

bottom of the Product: 

a)  Use a flat-blade screwdriver to push the tab in 

the recess of each foot and pull the foot from 
the cover. 

b)  Store the feet for future use. 

Содержание Y9600

Страница 1: ...o rack mount the Product To install the slides 1 Push a flat blade screwdriver between the chassis and the plastic strip to remove the plastic side trim from the Product Pry outward until the strip is free of the two grooves and then pull the strip free from the Product 2 Attach each rack mount ear to the front of the Product Note So that the bezel is not detached complete one side before you star...

Страница 2: ...2 inch truss head screws and two 10 24 U nuts See Figure 3 Detail B 2 Install the rear slide brackets longer pair to the equipment rack mounting brackets as described in step 1 3 Attach the slides to the front brackets with two 8 32 x 1 2 inch pan head screws two 8 flat washers two 8 lock washers and two 8 hex nuts See Figure 3 Align the access hole in the slide by moving the slide in and out as n...

Страница 3: ... Secure the Product front rack ears to the rack mounting rails with two 10 32 x inch decorative head screws per ear If your rack has unthreaded mounting rails install two 10 32 U nuts in the appropriate places on the rail holes Decorative Head 10 32 x 3 4 in 4 U Nut 10 32 4 Hex Nut 18 8 Flat Washer 8 8 Lock Washer 8 8 U Nut 10 24 8 Pan Head 8 32 x 1 2 in 12 Truss Head 8 32 x 1 4 in 10 Truss Head 1...

Страница 4: ...ack Contact Fluke Calibration To contact Fluke Calibration call one of the following telephone numbers Technical Support USA 1 877 355 3225 Calibration Repair USA 1 877 355 3225 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 40 2675 200 Japan 81 3 6714 3114 Singapore 65 6799 5566 China 86 400 810 3435 Brazil 55 11 3759 7600 Anywhere in the world 1 425 446 6110 To see product information and downlo...

Отзывы: