Precision Process Calibrator
Product Description
11
Item Name
Description
Selects Pressure input mode. Subsequent pushes of this key
cycle through the pressure units. Pressure input mode uses the
pressure module connector on the primary input/output side.
Each side can use pressure mode at the same time and can be
set to show the same pressure measurement in different units if
necessary.
Input
Terminals
[1,2]
Common input terminals for dc voltage and current.
[1]
These terminal binding posts are made of a special copper alloy to decrease thermal EMF's.
Discreet wires or standard banana plugs can be used, and the HI/LO pairs are distanced for standard
dual banana plugs.
[2]
Warning: To prevent possible electrical shock, fire, or personal injury Do not exceed a
maximum of 100 V peak to chassis ground.
Figure 4. Isolated Input Display, Controls, and Terminals (cont.)
Rear Panel
Figure 5 shows the rear-panel layout.
I
o
3
4
7
6
2
1
5
gwp05.eps
Item Description
RS-232 9-pin connector for remote control of the Product with a computer serial interface.
GPIB IEEE 488.2 connector for remote control of the Product with a GPIB bus.
Service port to update the Product firmware.
Chassis ground terminal internally connected to the ground prong of the AC power inlet.
Standard IEC AC power inlet for 120/240 V ac.
Main power on/off switch.
Power line voltage selector and fuse compartment. See “Maintenance” in Chapter 7 of the
Users Manual for instructions to change the line voltage selector and fuses.
Figure 5. Rear Panel
Содержание 7526A
Страница 4: ...7526A Getting Started ii ...
Страница 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Error Messages 15 3 Line Voltage Settings 15 ...
Страница 6: ...7526A Getting Started iv ...
Страница 8: ...7526A Getting Started vi ...