Reference Pressure Gauge
Battery Life
9
How to Set a Custom Engineering Unit or Scale
The last menu selection in Supervisory Mode is
SET FACTR
. You can set a multiplier factor from 0.001 to
100 to make a custom scale. The set factor is multiplied
by the psi measured and the result is shown.
Example: 40 psi is the equivalent of 1000 lbs of product in
a tank. It is necessary to show the product weight with a
100 psi gauge. If you set a factor of 25, 40 psi pressure
shows as 1000 (40 x 25). The engineering unit shown is
Cust
(custom).
Battery Life
Battery life is approximately 75 hours of operation with
the backlight off. When the battery voltage is low, the low-
battery icon (
) shows on the top left of the display. To
replace the batteries, see the How to Change the
Batteries section.
Maintenance
How to Clean the Product
Clean the Product with a soft cloth dampened with water
or water and weak soap.
W
Caution
To prevent possible damage to the Product,
do not use solvents or abrasive cleansers.
W
Caution
For safe operation and maintenance of the
product:
•
Repair the Product before use if the
battery leaks.
•
Remove batteries to prevent battery
leakage and damage to the Product if it
is not used for an extended period.
•
Be sure that the battery polarity is
correct to prevent battery leakage.
•
Have an approved technician repair the
Product.
www.
.com
1.800.561.8187