background image

 

Reference Pressure Gauge 

 

Operation 

engineering units. See the Supervisory Mode section for 
instructions to set up custom units. 

Set Auto Off 

Auto Off can be set in 1 minute increments from 1 to 30 
minutes or you can turn off the function for continuous 
Product operation. The Product is configured for 30 
minutes. Push 

 and 

 to set the necessary interval. 

The “off” position is at the low end of the selections, less 
than 1 minute. 

Show Battery Voltage 

Actual battery voltage and a percent-of-life bargraph 
show the battery charge. No adjustments are made in this 
parameter. 

Display Actual Temperature 

The Product is temperature compensated This parameter 
shows the temperature measured by the internal sensor. 
Push 

 or 

 to show degrees F or C.  

Set Damping 

Selections are “on” 

 and “off” 

. Damping smooths 

readings from pulsating pressure sources. 

Set Sample Rate 

This function finds how many times pressure is sampled 
and the display is updated. Selections are 0.5, 1, 3, and 
10 samples/second. Note that 10/second gives the fastest 
response time. 

Set Tare 

Use this function to set a constant offset value that is then 
subtracted from the measured pressure. For example, if a 
tare is set at 30 psi, and the measured pressure is 37 psi, 
7 psi is shown.   

A pressure of 27 psi is shown as -3 psi.  

Push 

 and 

 to set the tare value. The value is in 

relation to the engineering units and resolution selected 
for display. Tare value can be set to the maximum range 
of the gauge.  

For safety, the bar graph always shows the actual 
pressure in relation to the full range of the gauge 
regardless of the tare position. This is done to make sure 
that with a “0” reading pressure is being applied to the 
gauge.   

www.

.com

[email protected]

1.800.561.8187

Содержание 2700G Series

Страница 1: ...tion All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 2700G Series Reference Pressure Gauge Users Manual www com i...

Страница 2: ...repair or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period uke assumes no risk for damage in transit Following warranty repair the p...

Страница 3: ...ard Equipment 2 Safety Information 2 Special Conditions for Safe Use 3 Symbols 4 Display and Buttons 5 Operation 6 How to Setup the Product 6 Engineering Units 6 Set Auto Off 7 Show Battery Voltage 7...

Страница 4: ...an the Product 9 How to Change the Batteries 10 Accessories 11 USB Interface Cable 11 Power Module 11 Serial Interface Instructions 13 Initiating Communication 13 List of Commands 13 Parameter Units 1...

Страница 5: ...iii List of Tables Table Title Page 1 Symbols 4 2 Display and Buttons 5 3 Units Conversion 16 www com information itm com 1 800 561 8187...

Страница 6: ...v List of Figures Figure Title Page 1 The Product 5 2 Change the Batteries 10 3 USB Power Module and Universal Connectors 12 www com information itm com 1 800 561 8187...

Страница 7: ...ference or in applications where high accuracy pressure measurement is required The Product features user configurable functions that include Sampling rate Tare Damping Auto off Min Max When the Produ...

Страница 8: ...le and operate high pressure systems if you know the correct safety procedures High pressure liquids and gases are hazardous and the energy from them can be released without warning To prevent possibl...

Страница 9: ...torque specified Maximum torque specified is 20 Nm 15 ft lb Special Conditions for Safe Use Misuse If the Product is exposed to overpressure or sudden physical shock such as being dropped examine it f...

Страница 10: ...relevant Australian standards This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicates that you must not discard this electrical electronic product in...

Страница 11: ...uct Zeros the display In Configure Mode push the button to move forward through the menus MIN MAX records minimum and maximum pressure values and saves them to memory Push to show maximum MAX indicati...

Страница 12: ...ord a Tare value the pressure shown is not the actual pressure applied How to Setup the Product Before you use the product it is necessary to configure it for your application Push to go to the Setup...

Страница 13: ...tions are on and off Damping smooths readings from pulsating pressure sources Set Sample Rate This function finds how many times pressure is sampled and the display is updated Selections are 0 5 1 3 a...

Страница 14: ...word entry Hold or down to move faster through the selections by a factor of 10 When you stop the counter push and again to move forward or backward by a factor of 1 The password is factory set and ca...

Страница 15: ...eration with the backlight off When the battery voltage is low the low battery icon shows on the top left of the display To replace the batteries see the How to Change the Batteries section Maintenanc...

Страница 16: ...batteries see Figure 2 1 Pull off the Product cover 2 Use a Phillips screwdriver to loosen the captive screw on the battery door 3 Remove the battery door 4 Replace the three AA batteries 5 Install th...

Страница 17: ...is on the back of the Product You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC For specifications on the interface see the Sp...

Страница 18: ...2700G Series Users Manual 12 gsn003 eps Figure 3 USB Power Module and Universal Connectors www com information itm com 1 800 561 8187...

Страница 19: ...ent list of commands to communicate with the Gauge CLS Clears the error queue FAULT Returns an error code from the error queue IDN Identification query Returns the manufacturer model number and firmwa...

Страница 20: ...meter Units The subsequent list of units is used with the Product psi Pressure in pounds per square inch bar Pressure in bars mBar Pressure in millibars kg cm2 Pressure in kilograms per centimeter squ...

Страница 21: ...s The gauge uses the subsequent error codes 101 A non numeric entry was received where it should be a numeric entry 102 Too many significant digits entered 103 Invalid units or parameter value receive...

Страница 22: ...tor from kPa Conversion Factor to kPa psi 0 14503773773 6 894759 bar 0 01 100 MPa 0 001 1000 kgf cm2 0 010197162130 98 06652 inH2O 4 C 4 014742 0 249082 inH2O 20 C 4 021845 0 248642 inH2O 60 F 4 01858...

Страница 23: ...Factor to kPa cmH2O 4 C 10 19744 0 09806383 cmH2O 20 C 10 21549 0 09789056 mH2O 4 C 0 1019744 9 806383 mH2O 20 C 0 1021549 9 789056 kPa 1 1 mbar 10 0 1 inHg 0 C 0 2952998 3 386387 mmHg 0 C 7 500618 0...

Страница 24: ...an dry non corrosive gas 100 300 500 1000 psi any liquids or gases compatible with 316 stainless steel Above 1000 psi any non flammable non toxic non explosive non oxidizing liquid or gas compatible w...

Страница 25: ...5 x 5 in depth 1 5 in Without Protective Cover Pressure Connection in NPT Male Housing Cast ZNAL Display 5 1 2 Digits 16 53 mm 0 65 in high 20 Segment bar graph 0 to 100 Power Battery three size AA a...

Страница 26: ...3000 5000 10000 Pressure Range MPa 0 1 0 2 0 7 2 3 5 7 20 35 70 Vacuum Range psi 15 15 12 12 12 12 0 0 0 Vacuum Range kPa 100 100 80 80 80 80 0 0 0 Burst Pressure psi 45 90 1000 2000 2000 10000 10000...

Отзывы: