FLOW UNIT
XS
S
M
L
XL
Sensor
inner
diameter
25 µm
150 µm
430 µm
1.0 mm
1.8 mm
Maximum
pressure
200 bar
200 bar
100 bar
12 bar
5 bar
Wetted
materials
PEEK and Quartz
Glass
PEEK and Quartz
Glass
PEEK and
Borosilicate Glass
PEEK and
Borosilicate Glass
PEEK and
Borosilicate
Glass
Calibrated
Media
Water
Water
Isopropyl Alcohol
Water
Isopropyl Alcohol
Water
Isopropyl
Alcohol
Water
Range
0±1.5 µL/min
0±7 µL/min
0±70 µL/min
0±80 µL/min
0±500 µL/min
0±1 mL/min
0±10 mL/min
0±5mL/min
Accuracy
(m.v =
measured
value)
10% m.v. above
75 nL/min
5% m.v.above 0.42
µL/min
20% m.v. above 4.2
µL/min
5% m.v. above 2.4
µL/min
20% m.v above 25
µL/min
5% m.v. above
0.04 mL/min
20% m.v above 0.5
mL/min
5% m.v.above
0.2 mL/min
7.5 nL/min below
75 nL/min
21 nL/min below
0.42 µL/min
210 nL/min below
4.2 µL/min
0.12 µL/min below
2.4 µL/min
5 µL/min below 25
µL/min
1.5 µL/min below
0.04 mL/min
100 µL /min below
0.5 mL/min
10 µL/min below
0.2 mL/min
Lowest
detectable flow
increment
3.7 nL/min
10 nL/min
0.06 µL/min
0.7 µL/min
3 µL/min
Warning:
Please note that the maximum pressure depends on the FLOW UNIT model. Ensure that the
pressure applied to a FLOW UNIT does not go beyond this value at all times.
The Flow-Rate Platform suits your own fluid controller. If you use a pressure regulator you may have to enter
a maximum pressure below this value. If you use other flow controller, be aware that pressure may go
higher than 100 bar very easily and may cause damage to your FLOW UNIT.